The rebreathing characteristics of the Hudson oxygen mask.

N T Campkin1, R G Ooi, N C Soni.   

Abstract

A lung model was used to study the effect of varying minute ventilation, oxygen flow and inspiratory : expiratory ratio on the rebreathing characteristics of the Hudson oxygen mask. The onset of rebreathing was detected using two different sets of criteria; those defined by Kain et al. and Humphrey. The onset of rebreathing was also quantified in terms of the inspired volume of carbon dioxide and increases in deadspace. The magnitude of rebreathing was increased by reducing oxygen inflow and increasing minute ventilation. Variation in I : E ratio between 1 : 1 and 1 : 4 had little effect on rebreathing. Using the clinical criteria of Kain et al, the onset of rebreathing was detected when oxygen inflow/minute ventilation ratio was reduced below 0.78 on average. With the other criterion, rebreathing occurred when the ratio was 0.46.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8460805     DOI: 10.1111/j.1365-2044.1993.tb06911.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  2 in total

1.  Capnographic Analysis of Minimum Mandatory Flow Rate for Hudson Face Mask: A Randomized Double-blind Study.

Authors:  Pranjali Kurhekar; T Krishna Prasad; Buddhan Rajarathinam; M S Raghuraman
Journal:  Anesth Essays Res       Date:  2017 Apr-Jun

2.  Rapid reduction of macular edema due to retinal vein occlusion with low-dose normobaric hyperoxia.

Authors:  Jorge G Arroyo; Brendan Seto; Keiko Yamada; Ke Zeng; Robert Minturn; Colin A Lemire
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-02-22       Impact factor: 3.117

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.