Flow velocity profile of the pulmonary artery measured by the continuous cardiac output monitoring catheter.

K Miyasaka1, M Takata, K Miyasaka1.   

Abstract

The KATS catheter (continuous arterial thermodeprivation system catheter) measures the blood flow velocity of the pulmonary artery (PA) by thermodeprivation which enables continuous determination of cardiac output. The accuracy of this system may depend on the degree of uniformity of flow velocity in the PA, because small movements of the catheter within the PA are inevitable with a beating heart. We evaluated the flow velocity profile of the PA in seven anaesthetized open-chest dogs to assess these potential errors. A custom-made stiff catheter, at the tip of which was incorporated the flow velocity sensor of the KATS catheter, was used to penetrate the main PA in the short axis direction (perpendicular to flow direction) or the long axis direction (along flow direction). The stiff catheter was moved in increments of 2.5 mm, and flow velocity was recorded. The wall-to-wall distance of the PA along each direction was divided into five sections (S1 to S5 for the short axis, and L1 to L5 for the long axis). Flow velocity data for each section were averaged and presented as relative values against the control mid-point velocity. Along the short axis, flow velocity was 0.41 +/- 0.20 (SD), 1.00 +/- 0.10, 1.03 +/- 0.10, 1.08 +/- 0.13 and 0.49 +/- 0.26 from S1 to S5, i.e., lower in S1 and S5 which were close to the vascular walls (P < 0.05) but uniform in other areas. Along the long axis, flow velocity was 0.28 +/- 0.28, 0.88 +/- 0.09, 0.94 +/- 0.08, 1.06 +/- 0.25 and 1.28 +/- 0.50 from L1 to L5.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Year:  1993        PMID: 8443859     DOI: 10.1007/BF03011318

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  6 in total

Review 1.  Current concepts measurement of cardiac output by thermodilution.

Authors:  R D Weisel; R L Berger; H B Hechtman
Journal:  N Engl J Med       Date:  1975-03-27       Impact factor: 91.245

2.  Continuous cardiac output determination by thermodeprivation.

Authors:  K Miyasaka; G Volgyesi; M Katayama; S Tanabe
Journal:  J Anesth       Date:  1991-07       Impact factor: 2.078

3.  A new technique for measurement of cardiac output by thermodilution in man.

Authors:  W Ganz; R Donoso; H S Marcus; J S Forrester; H J Swan
Journal:  Am J Cardiol       Date:  1971-04       Impact factor: 2.778

4.  Catheterization of the heart in man with use of a flow-directed balloon-tipped catheter.

Authors:  H J Swan; W Ganz; J Forrester; H Marcus; G Diamond; D Chonette
Journal:  N Engl J Med       Date:  1970-08-27       Impact factor: 91.245

5.  Velocity profiles in the main pulmonary artery of dogs and man, measured with a thin-film resistance anemometer.

Authors:  S R Reuben; J P Swadling; G J de Lee
Journal:  Circ Res       Date:  1970-12       Impact factor: 17.367

6.  The hot-film anemometer--a method for blood velocity determination. II. In vivo comparison with the electromagnetic blood flowmeter.

Authors:  P K Paulsen
Journal:  Eur Surg Res       Date:  1980       Impact factor: 1.745

  6 in total
  4 in total

1.  Analysis of the accuracy of continuous thermodilution cardiac output measurement. Comparison with intermittent thermodilution and Fick cardiac output measurement.

Authors:  L Jacquet; G Hanique; D Glorieux; P Matte; M Goenen
Journal:  Intensive Care Med       Date:  1996-10       Impact factor: 17.440

2.  Continuous cardiac output: myth or reality?

Authors:  J G Ramsay
Journal:  Can J Anaesth       Date:  1993-02       Impact factor: 5.063

3.  Movement of pulmonary artery catheters.

Authors:  M S Chew; E Sloth; M Veien; K V Jensen; H H Kure; J M Hasenkam
Journal:  Heart Vessels       Date:  1996       Impact factor: 2.037

Review 4.  Considerations for numerical modeling of the pulmonary circulation--a review with a focus on pulmonary hypertension.

Authors:  V O Kheyfets; W O'Dell; T Smith; J J Reilly; E A Finol
Journal:  J Biomech Eng       Date:  2013-06       Impact factor: 2.097

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.