Design of a microwave system for endoscopy: an experimental study of energy, tissue contact, and hemostatic efficacy.

A A Kalabakas1, A J Porter, L Mule, M J Birch, D J Pollock, C P Swain.   

Abstract

BACKGROUND: A microwave generator and delivery system for endoscopic use was built. Using a 650-W, 2450-MHz magnetron, 0-160 W were generated from the tip of a 180-cm flexible coaxial cable (2.1 mm diameter).
METHODS: Three methods of achieving hemostasis with microwaves were identified studying standard bleeding canine ulcers: (1) interstitial method: inserting the coaxial tip into the tissue and heating slowly until bleeding stopped; (2) contact method: tip held in contact, light pressure applied; (3) noncontact method: microwave-induced sparking (dielectric breakdown) with tip held 1 mm from tissue.
RESULTS: Studies of optimal energy levels for hemostasis showed that high power (70 W) noncontact methods required significantly less energy to stop bleeding than contact or interstitial methods. The noncontact method was more effective than the contact method, stopping 20 of 20 bleeding ulcers vs. 10 of 20 (P < 0.001) and was more rapidly effective causing less tissue damage (P < 0.05) than the interstitial method.
CONCLUSIONS: In a randomized comparison using a non-contact method, microwave coagulation was superior (P < 0.001) to a polidocanol 1% + adrenaline 1:10,000 injection and control treatment stopping 40 of 40 vs. 0 of 20 and 0 of 20 standard bleeding ulcers. Microwaves stopped bleeding from 10 of 10 severed mesenteric vessels, whereas injection was ineffective (0 of 10, P < 0.001). Microwaves look promising for hemostasis at flexible endoscopy.

Entities:  

Mesh:

Year:  1993        PMID: 8440429     DOI: 10.1016/0016-5085(93)91002-y

Source DB:  PubMed          Journal:  Gastroenterology        ISSN: 0016-5085            Impact factor:   22.682


  2 in total

1.  A randomised controlled comparison of injection, thermal, and mechanical endoscopic methods of haemostasis on mesenteric vessels.

Authors:  C C Hepworth; S S Kadirkamanathan; F Gong; C P Swain
Journal:  Gut       Date:  1998-04       Impact factor: 23.059

2.  Microwave coagulation versus sclerotherapy after band ligation to prevent recurrence of high risk of bleeding esophageal varices in Child-Pugh's A and B patients.

Authors:  Leonardo Trevizan Monici; José Olympio Meirelles-Santos; Elza Cotrim Soares; Maria Aparecida Mesquita; José Murilo Robilotta Zeitune; Ciro Garcia Montes; Jazon Romilson Souza Almeida; Ademar Yamanaka; Luis Alberto Magna
Journal:  J Gastroenterol       Date:  2009-10-03       Impact factor: 7.527

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.