A new asymmetric screen-film combination for conventional chest radiography: evaluation in 50 patients.

S J Swensen1, J E Gray, L R Brown, G L Aughenbaugh, G F Harms, J Stears.   

Abstract

OBJECTIVE: A new asymmetric screen-film combination for conventional chest radiography was evaluated by four thoracic radiologists.
MATERIALS AND METHODS: Fifty chest radiographs obtained with a new asymmetric screen-film combination were compared with radiographs obtained in the same patients during the same breath-hold with a conventional system. Patients referred from the oncology service were studied because of the increased prevalence of abnormal findings on chest radiographs in this population. The differences in the quality of the images were assessed by four chest radiologists who used a direct-comparison technique. Eleven anatomic shadows that are normal on chest radiographs were assessed. Fifty abnormal findings also were compared.
RESULTS: The image quality of the new asymmetric screen-film combination was judged to be superior to that of the conventional system for normal anatomy and abnormal findings (p < or = .003).
CONCLUSION: The new asymmetric film consistently showed more clinically useful detail in the mediastinum and in the lung projected over the diaphragm and heart. The radiation dose was about 30% less with the new screen-film system.

Entities:  

Mesh:

Year:  1993        PMID: 8430540     DOI: 10.2214/ajr.160.3.8430540

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  3 in total

1.  Quality-of-service improvements from coupling a digital chest unit with integrated speech recognition, information, and picture archiving and communications systems.

Authors:  W Pavlicek; J R Muhm; J M Collins; B Zavalkovskiy; B S Peter; M D Hindal
Journal:  J Digit Imaging       Date:  1999-11       Impact factor: 4.056

2.  Evaluation of irreversible compression of digitized posterior-anterior chest radiographs.

Authors:  B J Erickson; A Manduca; K R Persons; F Earnest; T E Hartman; G F Harms; L R Brown
Journal:  J Digit Imaging       Date:  1997-08       Impact factor: 4.056

3.  Efficiency and reporting confidence analysis of sequential dual-energy subtraction for thoracic x-ray examinations.

Authors:  Mehmet Can Gezer; Oktay Algin; Aytac Durmaz; Halil Arslan
Journal:  Qatar Med J       Date:  2019-09-23
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.