Literature DB >> 842958

A rebreathing method for measuring carbon monoxide diffusing capacity. A supplement to the single-breath method.

R Marshall.   

Abstract

A rebreathing method for measuring carbon monoxide diffusing capacity is described. The method can be used in patients whose vital capacity is too small, or whose cooperation is inadequate, for the single-breath method. The apparatus is simple and the same gas mixtures are used as for the single-breath method. The results obtained by the rebreathing method are similar to those by the single-breath method (80% of the results by the rebreathing method lie within 20% of the value by the single-breath method). The rebreathing method also provides a measurement of the functional residual capacity.

Entities:  

Mesh:

Substances:

Year:  1977        PMID: 842958     DOI: 10.1164/arrd.1977.115.3.537

Source DB:  PubMed          Journal:  Am Rev Respir Dis        ISSN: 0003-0805


  2 in total

1.  Clinical assessment of a rebreathing method for measuring pulmonary gas transfer.

Authors:  N J Russell; L R Bagg; J Dobrzynski; D T Hughes
Journal:  Thorax       Date:  1983-03       Impact factor: 9.139

2.  Assessment of progression of asbestosis in the sheep model by bronchoalveolar lavage and pulmonary function tests.

Authors:  R Bégin; M Rola-Pleszczynski; S Massé; D Nadeau; G Drapeau
Journal:  Thorax       Date:  1983-06       Impact factor: 9.139