Literature DB >> 8426320

Calculation of the thickness of an intraocular lens.

K Naeser1, E V Naeser.   

Abstract

Lens manufacturers do not usually supply information about the central thickness of an intraocular lens optic. This paper describes a method to calculate the central lens thickness from variables normally supplied by the manufacturer, i.e., the total lens power in situ, the edge thickness, the optic diameter, and some knowledge of the optic architecture. The formulas are universal and valid for any lens, regardless of lens design and material. There is a correlation between calculated values of optic thickness and values supplied by manufacturers. Individually estimated lens thickness may increase the accuracy of intraocular lens power calculation, pseudophakic anterior chamber depth estimation, and pseudophakic axial length calculation.

Mesh:

Year:  1993        PMID: 8426320     DOI: 10.1016/s0886-3350(13)80278-7

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  1 in total

1.  In Vivo Intraocular Lens Thickness Measurement and Power Estimation Using Optical Coherence Tomography.

Authors:  Ehsan Barzanouni; Diba Idani; Farideh Sharifipour
Journal:  J Ophthalmic Vis Res       Date:  2022-08-15
  1 in total