Evaluation of a one-piece poly(methyl methacrylate) intraocular lens with a 7 mm biconvex optic and a total diameter of 10 mm.

M Amon1, R Menapace.   

Abstract

To find out how large the total diameter of an open-loop intraocular lens (IOL) should be, we studied 90 consecutive implantations of a one-piece poly(methyl methacrylate) IOL with a 7 mm biconvex optic and a total diameter of 10 mm (Adatomed 75 ST). The lens design made implantation simple, even through a small capsulorhexis. In 85% of the cases, the IOL remained well centered. One case with asymmetric placement of the IOL showed a decentration of more than 1 mm. In 14% of the eyes slight decentrations occurred because the capsular bag was too large. Contact between the iris and IOL was found in 15%. In 4% there were folds in the posterior capsule, and in 35% there was a small space between the IOL and the posterior capsule because the capsular bag was too large. Although the ST lens has a small total diameter, it remained at a distance from the iris in most cases. In-the-bag implantation through a continuous central capsulorhexis is mandatory to avoid secondary displacement; nevertheless, slight decentrations cannot be excluded. For a large capsular bag, the total diameter of an IOL should exceed 10 mm. Clinical results, cellular reaction on the lens surface (26.6%), biocompatibility, and visual performance are comparable to those of other well-approved poly(methyl methacrylate) IOLs.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8426314     DOI: 10.1016/s0886-3350(13)80273-8

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  1 in total

1.  Bilateral identical intervals between phacoemulsification procedures performed 23 years before retinal detachment.

Authors:  Chrysanthos Symeonidis; Lampros Lamprogiannis; Ioannis Tsinopoulos
Journal:  Oman J Ophthalmol       Date:  2018 Sep-Dec
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.