Literature DB >> 8424289

Verification of the position of a central venous catheter by intra-atrial ECG. When does this method fail?

L Salmela1, U Aromaa.   

Abstract

The purpose of this study was to determine why intra-atrial ECG tracing for checking the position of a central venous catheter fails in certain patients. Three hundred and fifty prospective and consecutive patients scheduled for central venous catheterization using various puncture sites and techniques were investigated. The catheters were 20 cm in length. After its introduction, the catheter was connected to an Alphacard (Sterimed, Saarbrücken) for the intra-atrial ECG tracing. The method failed in 29 patients, of whom nine had manifest myocardial pathology. In two patients the catheter looped, while in the remaining 18 the catheter proved to be too short. In these 18 patients, the cannulation was mainly performed via the external jugular vein and/or from the left side. Most of the patients were elderly males, and 11 of the 18 patients showed radiological signs of pulmonary emphysema. In such individuals it is advisable to use a catheter longer than 20 cm.

Entities:  

Mesh:

Year:  1993        PMID: 8424289     DOI: 10.1111/j.1399-6576.1993.tb03592.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  4 in total

1.  [Modified ECG-guidance for optimal central venous catheter tip positioning. A transesophageal echocardiography controlled study].

Authors:  W Schummer; C Schummer; C Schelenz; P Schmidt; R Fröber; E Hüttemann
Journal:  Anaesthesist       Date:  2005-10       Impact factor: 1.041

Review 2.  A review of calcium channel antagonists in the treatment of pediatric hypertension.

Authors:  Shobha Sahney
Journal:  Paediatr Drugs       Date:  2006       Impact factor: 3.022

3.  External jugular vein catheterization using 'intra-atrial electrocardiogram'.

Authors:  Dilek Karaaslan; Ugur Altinisik; Tulay Tuncer Peker; Esra Nayir; Sadik Ozmen
Journal:  Yonsei Med J       Date:  2009-04-30       Impact factor: 2.759

4.  Depth of central venous catheterization by intracardiac electrocardiogram in adults.

Authors:  Prerana N Shah; Deepa Kane; Jithesh Appukutty
Journal:  Anesth Pain Med       Date:  2013-01-01