Literature DB >> 8419573

Torque and pullout analysis of six currently available self-tapping and "emergency" screws.

J M Boyle1, D E Frost, W L Foley, J J Grady.   

Abstract

Insertional torque (IT), stripping torque (ST), and uniaxial pullout tests were performed to evaluate the effectiveness of six screw systems (Wurzburg, Techmedica, Synthes, Timesh, Steinhouser, Luhr) in thin porcine rib. The Timesh 2.2-mm self-tapping screw produced the largest insertional and stripping torque of all systems tested as well as the largest difference between the insertional and stripping torque. The Timesh emergency screw also had the largest insertional and stripping torque as well as the largest difference between the insertional and stripping torque. In pullout tests, the Timesh screw was found to be the most retentive. The overall data indicated that the ideal self-tapping screw should have the largest difference possible between drill size and external diameter, a channel, and at least three self-tapping threads for maximum retention.

Entities:  

Mesh:

Year:  1993        PMID: 8419573     DOI: 10.1016/s0278-2391(10)80389-1

Source DB:  PubMed          Journal:  J Oral Maxillofac Surg        ISSN: 0278-2391            Impact factor:   1.895


  2 in total

1.  Experimental and clinical evaluation of laryngeal chondrosynthesis.

Authors:  R Siegert; E Zimmermann; P Oppermann; S Remmert; K H Ahrens; H Weerda
Journal:  Eur Arch Otorhinolaryngol       Date:  1996       Impact factor: 2.503

2.  Biomechanical evaluation of peak reverse torque (PRT) in a dynamic compression plate-screw construct used in a goat tibia segmental defect model.

Authors:  Remigiusz M Grzeskowiak; Carrie Wheeler; Elizabeth Taylor; James Lillich; James Roush; Alexandru S Biris; David E Anderson
Journal:  BMC Vet Res       Date:  2019-09-05       Impact factor: 2.741

  2 in total