Failure of the porous-coated anatomic prosthesis in total knee arthroplasty due to severe polyethylene wear.

A Tsao1, L Mintz, C R McRae, S D Stulberg, T Wright.   

Abstract

Four hundred and eighty-seven total knee arthroplasties were performed by a single surgeon with use of a porous-coated anatomic prosthesis between 1982 and 1989. There were thirty-two clinical failures (7 per cent) due to severe wear of the surfaces of the tibial and patellar polyethylene components. Thirty patients had a revision. The average time to failure of the implant was four and one-half years. The initial clinical symptoms of failure by wear consisted of a painless effusion with a decreased range of motion. Subsequent pain was considered as the criterion for failure necessitating operative intervention. Increased weight and decreased age of the patient and a thinner tibial component were significant predictors of an increased risk of failure (p < 0.01). Examination of retrieved tibial components revealed extensive delamination caused by fracture of the polyethylene at a depth of about one millimeter below the surface. Cracks that had propagated in from the medial and lateral peripheries of the tibial component toward the center of the condyles were also a common finding. It appears that the design of the implant as well as clinical factors (the age and weight of the patient) contributed to the mechanical failure of the polyethylene of these implants.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8419386     DOI: 10.2106/00004623-199301000-00004

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  3 in total

1.  Polyethylene thickness is a risk factor for wear necessitating insert exchange.

Authors:  Bart G Pijls; Henrica M J Van der Linden-Van der Zwaag; Rob G H H Nelissen
Journal:  Int Orthop       Date:  2011-11-22       Impact factor: 3.075

2.  Computer-assisted design of the sagittal shapes of a ligament-compatible total ankle replacement.

Authors:  A Leardini; F Catani; S Giannini; J J O'Connor
Journal:  Med Biol Eng Comput       Date:  2001-03       Impact factor: 3.079

3.  Patient and surgery related factors associated with fatigue type polyethylene wear on 49 PCA and DURACON retrievals at autopsy and revision.

Authors:  Markus Rohrbach; Martin Lüem; Peter E Ochsner
Journal:  J Orthop Surg Res       Date:  2008-02-22       Impact factor: 2.359

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.