Literature DB >> 8403287

Transcatheter occlusion of patent ductus arteriosus with Gianturco coils.

T R Lloyd1, R Fedderly, A M Mendelsohn, S K Sandhu, R H Beekman.   

Abstract

BACKGROUND: Transcatheter occlusion with Gianturco coils has been attempted in a small number of patients with tiny (< or = 1.5-mm diameter) patent ductus arteriosus, and preliminary results have been encouraging. The present study extends this method to larger ductus sizes and makes recommendations for proper coil size selection. METHODS AND
RESULTS: Coil occlusion was attempted in 24 consecutive patients with patent ductus arteriosus who did not require other cardiac surgery. Median patient age was 4.2 years (8 months to 30 years), and mean ductus diameter was 1.7 +/- 0.8 mm. Two instances of coil embolization occurred in the first 4 patients, with successful coil retrieval. Based on this experience, we proposed that the coil helical diameter should be twice or more the minimum ductus diameter, with coil length sufficient for three or more loops. With these recommendations, coils were successfully implanted in the subsequent 20 consecutive patients. Of the 22 patients with successful coil implantation, 15 (68%) had no residual shunting, and 7 had trace residual shunting by angiography. The continuous murmur was abolished in all 22 patients. No significant complications occurred, and all patients were discharged within 24 hours of successful coil implantation. No change in the systolic pressure gradient between main and left pulmonary artery or ascending and descending aorta was observed.
CONCLUSIONS: Transcatheter occlusion of patent ductus arteriosus can be safely and effectively achieved in patients with ductus diameters up to 3.3 mm. Coil occlusion does not cause obstruction to flow in the left pulmonary artery or descending aorta. Coils should be selected to provide a helical diameter twice or more the minimum ductus diameter and a length sufficient for three or more loops.

Entities:  

Mesh:

Year:  1993        PMID: 8403287     DOI: 10.1161/01.cir.88.4.1412

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  23 in total

1.  Patent Ductus Arteriosus.

Authors: 
Journal:  Curr Treat Options Cardiovasc Med       Date:  1999-12

2.  Interventional pediatric cardiology: device closures.

Authors:  J L Wilkinson
Journal:  Indian J Pediatr       Date:  2000-07       Impact factor: 1.967

3.  [Percutaneous transcatheter coil embolization of the patent ductus arteriosus for elderly patient with left ventricular disfunction].

Authors:  S Watanabe; H Saitou; M Hata; M Miura; M Zuguchi; K Tabayashi
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  1998-07

4.  Occlusion of the neonatal patent ductus arteriosus with a simple retrievable device: a feasibility study.

Authors:  M B Neuss; J Y Coe; F Tio; T P Le; R Grabitz; D A Redel
Journal:  Cardiovasc Intervent Radiol       Date:  1996 May-Jun       Impact factor: 2.740

5.  Evolving use of embolisation coils for occlusion of the arterial duct.

Authors:  E Rosenthal; S A Qureshi; J Reidy; E J Baker; M Tynan
Journal:  Heart       Date:  1996-12       Impact factor: 5.994

6.  Non-surgical closure of large ductus arteriosus using Amplatzer Duct Occluder feasibility and early follow-up results.

Authors:  S Radhakrishnan; A Marwah; S Shrivastava
Journal:  Indian J Pediatr       Date:  2001-01       Impact factor: 1.967

Review 7.  Non surgical treatment in congenital heart disease.

Authors:  S Radhakrishnan; S Shrivastava
Journal:  Indian J Pediatr       Date:  1998 Mar-Apr       Impact factor: 1.967

8.  Amplatzer Device closure of Atrial Septal Defects and Patent Ductus Arteriosus: Initial Experience.

Authors:  J S Dugal; V Jetley; Charanjit Singh; S K Datta; J S Sabharwal; Sunil Sofat
Journal:  Med J Armed Forces India       Date:  2011-07-21

9.  Early implantation of multiple spring coils for severe haemolysis after incomplete transcatheter occlusion of persistent arterial duct.

Authors:  Y F Cheung; M P Leung; K T Chau
Journal:  Heart       Date:  1997-05       Impact factor: 5.994

10.  The results of transcatheter occlusion of patent ductus arteriosus: success rate and complications over 12 years in a single center.

Authors:  Deok Young Choi; Na Yeon Kim; Mi Jin Jung; Seong Ho Kim
Journal:  Korean Circ J       Date:  2010-05-27       Impact factor: 3.243

View more