T W Bauer1, B N Stulberg, J Ming, R G Geesink.
Abstract
Histologic sections of five hydroxyapatite-coated acetabular components retrieved at autopsy (2 dual geometry and 3 threaded cup designs) were analyzed, and the extent and pattern of bone apposition were compared with that of an uncoated, porous (beaded) dual geometry acetabular cup that had been removed for repeated dislocations. The results show hydroxyapatite and bone apposition most prominent in areas of likely load transmission, primarily around the peripheral rim of the dual-geometry cups and at the peaks of the threads of the screw cups. Hydroxyapatite and bone may have been removed by remodeling from the grooves between the threads. Although no significant wear debris was identified, the screw holes were incompletely covered by bone. The dual-geometry designs generally showed somewhat more bone apposition than the threaded cups, but all implants demonstrated less bone than might have been predicted from radiographs.Entities:
Mesh:
Substances:
Year: 1993 PMID: 8386747 DOI: 10.1016/s0883-5403(09)80010-7
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757