Abstract
A validated high-performance liquid chromatographic procedure employing ultraviolet detection for the analysis of diclofenac in human plasma is reported. The method is rapid and, coupled with column switching, leads to a sensitive, accurate and reproducible assay. The retention times of diclofenac and the internal standard (4'-methoxydiclofenac, CGP-4287) are 6.4 and 7.6 min, respectively. The peak height versus plasma concentration is linear over the range 5.0-2000 ng/ml with a detection limit below 2.5 ng/ml. The mean absolute recovery of diclofenac using the described assay is 96.5% (n = 24). The inter- and intra-day accuracy and precision are within 8.3% of the actual values for all concentrations investigated. Furthermore, this procedure is applied to assess the pharmacokinetics of a single 75-mg oral dose of diclofenac sodium.Entities:
Mesh:
Substances:
Year: 1993 PMID: 8376509 DOI: 10.1016/0378-4347(93)80396-l
Source DB: PubMed Journal: J Chromatogr