Using the laryngeal indices caliper to predict difficulty of laryngoscopy with a Macintosh #3 laryngoscope.

J T Roberts1, H H Ali, G D Shorten.   

Abstract

STUDY
OBJECTIVE: (1) To evaluate a device of the authors' design, the laryngeal indices caliper, which quantitates the position of the anterior edges of the larynx relative to the upper teeth and the external auditory canals; (2) to determine how relative laryngeal position affects ease of direct laryngoscopy with a Macintosh #3 laryngoscope.
DESIGN: Randomized, double-blind study.
SETTING: Inpatient surgery center at a university medical center. PATIENTS: 101 renal lithotripter patients.
INTERVENTIONS: Patients were measured with the laryngeal indices caliper prior to induction of general endotracheal anesthesia. They were then given a sleep dose of thiopental sodium (4 mg/kg) and paralyzed with a bolus dose of succinylcholine (1 mg/kg).
MEASUREMENTS AND MAIN RESULTS: Of the measurements taken or calculated, only laryngeal tilt (LT) showed a significant correlation with grade of difficulty of laryngoscopy. When the anterior surface of the thyroid cartilage was tilted more than 20 degrees anteriorly to a line perpendicular to the laryngeal indices line, the vocal cords could not be seen in 83% of the patients.
CONCLUSIONS: (1) Laryngeal tilt is a good predictor of difficulty of laryngoscopy with a Macintosh #3 laryngoscope; (2) the laryngeal indices caliper is a simple pocket device to measure LT indirectly.

Entities:  

Mesh:

Year:  1993        PMID: 8373608     DOI: 10.1016/0952-8180(93)90123-v

Source DB:  PubMed          Journal:  J Clin Anesth        ISSN: 0952-8180            Impact factor:   9.452


  2 in total

Review 1.  [Management of the upper airway in spontaneously breathing children. A challenge for the anaesthetist].

Authors:  B S von Ungern-Sternberg; T O Erb; F J Frei
Journal:  Anaesthesist       Date:  2006-02       Impact factor: 1.041

2.  Tracheal intubation and laryngoscope design.

Authors:  J W McIntyre
Journal:  Can J Anaesth       Date:  1993-03       Impact factor: 5.063

  2 in total