Design of a californium-based epithermal neutron beam for neutron capture therapy.

J C Yanch1, J K Kim, M J Wilson.   

Abstract

The potential of the spontaneously fissioning isotope, 252Cf, to provide epithermal neutrons for use in boron neutron capture therapy (BNCT) has been investigated using Monte Carlo simulation. The Monte Carlo code MCNP was used to design an assembly composed of a 26 cm long, 11 cm radius cylindrical D2O moderator followed by a 64 cm long Al filter. Lithium filters are placed between the moderator and the filter and between the Al and the patient. A reflector surrounding the moderator/filter assembly is required in order to maintain adequate therapy flux at the patient position. An ellipsoidal phantom composed of skull- and brain-equivalent material was used to determine the dosimetric effect of this beam. It was found that both advantage depths and advantage ratios compare very favourably with reactor and accelerator epithermal neutron sources. The dose rate obtainable, on the other hand, is 4.1 RBE cGy min-1, based on a very large (1.0 g) source of 252Cf. This dose rate is two to five times lower than those provided by existing reactor beams and can be viewed as a drawback of using 252Cf as a neutron source. Radioisotope sources, however, do offer the advantage of in-hospital installation.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8367525     DOI: 10.1088/0031-9155/38/8/013

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  1 in total

Review 1.  Physical, dosimetric and clinical aspects and delivery systems in neutron capture therapy.

Authors:  Bagher Farhood; Hadi Samadian; Mahdi Ghorbani; Seyed Salman Zakariaee; Courtney Knaup
Journal:  Rep Pract Oncol Radiother       Date:  2018-08-01
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.