Literature DB >> 8353887

Effects of active compression-decompression resuscitation on myocardial and cerebral blood flow in pigs.

K H Lindner1, E G Pfenninger, K G Lurie, W Schürmann, I M Lindner, F W Ahnefeld.   

Abstract

BACKGROUND: This study was designed to assess the effects of a modified cardiopulmonary resuscitation (CPR) technique that consists of both active compression and active decompression of the chest (ACD CPR) versus standard CPR (STD CPR) on myocardial and cerebral blood flow during ventricular fibrillation both before and after epinephrine administration. METHODS AND
RESULTS: During a 30-second period of ventricular fibrillation cardiac arrest, 14 pigs were randomized to receive either STD CPR (n = 7) or ACD CPR (n = 7). Both STD and ACD CPR were performed using an automated pneumatic piston device applied midsternum, designed to provide either active chest compression (1.5 to 2.0 in.) and decompression or only active compression of the chest at 80 compressions per minute and 50% duty cycle. Using radiolabeled microspheres, median total myocardial blood flow after 5 minutes of ventricular fibrillation was 14 (7 to 30, minimum to maximum) STD CPR versus 30 (9 to 46) mL.min-1 x 100 g-1 with ACD CPR (P < .05). Median cerebral blood flow was 15 (10 to 26) mL.min-1 x 100 g-1 with STD CPR and 30 (21 to 39) with ACD CPR (P < .01). When comparing STD with ACD CPR, aortic systolic (62 mm Hg [48 to 70] vs 80 [59 to 86]) and diastolic (22 [18 to 28] vs 28 [21 to 36]) pressures, calculated coronary systolic (30 [22 to 36] vs 49 [37 to 56]) and diastolic (18 [16 to 23] vs 26 [21 to 31]) perfusion pressures, end-tidal CO2 (1.4% [0.8 to 1.8] vs 2.1 (1.8 to 2.4]), cerebral O2 delivery (3.1 mL.min-1 x 100 g-1 [1.5 to 4.5] vs 5.3 [3.8 to 7.5]), and cerebral perfusion pressure (14 mm Hg [4 to 22] vs 26 [6 to 34]) were all significantly higher with ACD CPR: To compare these parameters before and after vasopressor therapy, a bolus of high-dose epinephrine (0.2 mg/kg) was given to all animals after 5 minutes of ventricular fibrillation. Organ blood flow and calculated perfusion pressures increased significantly in both the STD and ACD groups after epinephrine. The differences observed between STD and ACD CPR before epinephrine were diminished 90 seconds after epinephrine but were again statistically significant when assessed 5 minutes later, once the acute effects of epinephrine had decreased. No difference in short-term resuscitation success was found between the two groups.
CONCLUSIONS: We conclude that ACD CPR significantly increases myocardial and cerebral blood flow during cardiac arrest in the absence of vasopressor therapy compared with STD CPR:

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8353887     DOI: 10.1161/01.cir.88.3.1254

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  11 in total

1.  [Active compression-decompression resuscitation. Improved survival rate in an emergency medicine system with emergency physician assistance?].

Authors:  D Mauer; T Schneider; W Dick; D Elich; M Mauer
Journal:  Med Klin (Munich)       Date:  1997-07-15

2.  Cardiopulmonary resuscitation with active compression-decompression.

Authors:  C F Weston
Journal:  Br Heart J       Date:  1995-09

Review 3.  [Mechanical resuscitation assist devices].

Authors:  M Fischer; M Breil; M Ihli; M Messelken; S Rauch; J-C Schewe
Journal:  Anaesthesist       Date:  2014-03       Impact factor: 1.041

4.  Enhanced perfusion during advanced life support improves survival with favorable neurologic function in a porcine model of refractory cardiac arrest.

Authors:  Guillaume Debaty; Anja Metzger; Jennifer Rees; Scott McKnite; Laura Puertas; Demetris Yannopoulos; Keith Lurie
Journal:  Crit Care Med       Date:  2015-05       Impact factor: 7.598

Review 5. 

Authors:  J P Nolan; C D Deakin; J Soar; B W Böttiger; G Smith; M Baubin; B Dirks; V Wenzel
Journal:  Notf Rett Med       Date:  2006-02-01       Impact factor: 0.826

Review 6.  Drug administration in animal studies of cardiac arrest does not reflect human clinical experience.

Authors:  Joshua C Reynolds; Jon C Rittenberger; James J Menegazzi
Journal:  Resuscitation       Date:  2007-03-13       Impact factor: 5.262

7.  Improved chest recoil using an adhesive glove device for active compression-decompression CPR in a pediatric manikin model.

Authors:  Jai P Udassi; Sharda Udassi; Melissa A Lamb; Kenneth E Lamb; Douglas W Theriaque; Jonathan J Shuster; Arno L Zaritsky; Ikram U Haque
Journal:  Resuscitation       Date:  2009-08-15       Impact factor: 5.262

Review 8.  Cardiothoracic injuries after CardioPump CPR: a report of two cases and review of the literature.

Authors:  Martin Kolopp; Angélique Franchi; Patrick Grafiadis; Laurent Martrille
Journal:  Int J Legal Med       Date:  2017-09-29       Impact factor: 2.686

Review 9.  The Development of Innovative Handheld Devices to Augment Cardiopulmonary Resuscitation Therapy and External Cardioversion and Defibrillation.

Authors:  Melanie L Gershman; Brandon S Needelman; Sam N Schwarzwald; Todd J Cohen
Journal:  J Innov Card Rhythm Manag       Date:  2017-12-15

10.  Haemodynamic outcomes during piston-based mechanical CPR with or without active decompression in a porcine model of cardiac arrest.

Authors:  Mikkel T Steinberg; Jan-Aage Olsen; Morten Eriksen; Andres Neset; Per Andreas Norseng; Jo Kramer-Johansen; Bjarne Madsen Hardig; Lars Wik
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2018-04-24       Impact factor: 2.953

View more