Literature DB >> 8350832

Dose-volume histogram computations for small intracranial volumes.

H M Kooy1, L A Nedzi, E Alexander, J S Loeffler, R J Ledoux.   

Abstract

A sampling formalism is presented to accurately compute the absolute volumes and integral dose-volume histograms of small volumes treated in stereotactic radiosurgery. The presence of small volumes and sharp dose gradients places special constraints on the computational formalism and the accuracy required to compute the dose-volume relationships. We use a spatially nonuniform random sampling method to allow an efficient and accurate computation of the dose-volume histograms for an arbitrary number of volumes. The computation of absolute volume vs dose allows intercomparison of dose delivered to target and dose-critical volumes and allows a quantitative trade-off analysis often critical to an optimal treatment of the lesion.

Mesh:

Year:  1993        PMID: 8350832     DOI: 10.1118/1.597029

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  A simple plan evaluation index based on the dose to critical structures in radiotherapy.

Authors:  Ramachandran Prabhakar; Goura K Rath
Journal:  J Med Phys       Date:  2011-10
  1 in total