Literature DB >> 8350823

A technique for the evaluation of a missing tissue compensator system.

T Chu1, K Lee, P Dunscombe.   

Abstract

A two dimensional approach to the evaluation of a missing tissue compensator system has been developed and implemented. Quantitative descriptors of the dose area histograms generated by scanning films irradiated under a variety of compensated and uncompensated conditions are proposed. The method has been found to be useful during the acceptance testing of a commercial compensator system and the results of this exercise are presented.

Mesh:

Year:  1993        PMID: 8350823     DOI: 10.1118/1.597020

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Use of a matchline dosimetry analysis tool (MDAT) to quantify dose homogeneity in the region between abutting tangential and supraclavicular radiation fields.

Authors:  Kenneth L Homann; B Earl Gates; Mohammad Salehpour; David S Followill; Steven M Kirsner; R Allen White; Thomas A Buchholz; Karl L Prado
Journal:  J Appl Clin Med Phys       Date:  2010-07-27       Impact factor: 2.102