Literature DB >> 8346610

Deconvolution of in-vivo ultrasound B-mode images.

J A Jensen1, J Mathorne, T Gravesen, B Stage.   

Abstract

An algorithm for deconvolution of medical ultrasound images is presented. The procedure involves estimation of the basic one-dimensional ultrasound pulse, determining the ratio of the covariance of the noise to the covariance of the reflection signal, and finally deconvolution of the rf signal from the transducer. Using pulse and covariance estimators makes the approach self-calibrating, as all parameters for the procedure are estimated from the patient under investigation. An example of use on a clinical, in-vivo image is given. A 2 x 2 cm region of the portal vein in a liver is deconvolved. An increase in axial resolution by a factor of 2.4 is obtained. The procedure can also be applied to whole images, when it is ensured that the rf signal is properly measured. A method for doing that is outlined.

Entities:  

Mesh:

Year:  1993        PMID: 8346610     DOI: 10.1177/016173469301500204

Source DB:  PubMed          Journal:  Ultrason Imaging        ISSN: 0161-7346            Impact factor:   1.578


  6 in total

1.  Reflection-mode submicron-resolution in vivo photoacoustic microscopy.

Authors:  Chi Zhang; Konstantin Maslov; Song Hu; Ruimin Chen; Qifa Zhou; K Kirk Shung; Lihong V Wang
Journal:  J Biomed Opt       Date:  2012-02       Impact factor: 3.170

2.  Imaging with unfocused regions of focused ultrasound beams.

Authors:  Roger Zemp; Michael F Insana
Journal:  J Acoust Soc Am       Date:  2007-03       Impact factor: 1.840

3.  Blind deconvolution of medical ultrasound images: a parametric inverse filtering approach.

Authors:  Oleg Michailovich; Allen Tannenbaum
Journal:  IEEE Trans Image Process       Date:  2007-12       Impact factor: 10.856

4.  In vivo photoacoustic microscopy with 7.6-µm axial resolution using a commercial 125-MHz ultrasonic transducer.

Authors:  Chi Zhang; Konstantin Maslov; Junjie Yao; Lihong V Wang
Journal:  J Biomed Opt       Date:  2012-11       Impact factor: 3.170

5.  Ultrasound Open Platforms for Next-Generation Imaging Technique Development.

Authors:  Enrico Boni; Alfred C H Yu; Steven Freear; Jorgen Arendt Jensen; Piero Tortoli
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2018-07       Impact factor: 2.725

6.  Improved Ultrasound Microvessel Imaging Using Deconvolution with Total Variation Regularization.

Authors:  U-Wai Lok; Joshua D Trzasko; Chengwu Huang; Shanshan Tang; Ping Gong; Yohan Kim; Fabrice Lucien; Matthew R Lowerison; Pengfei Song; Shigao Chen
Journal:  Ultrasound Med Biol       Date:  2021-01-16       Impact factor: 2.998

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.