| Literature DB >> 8339884 |
J Germanaud1, J L Legoux, R Sabattier, X Causse, D H Trinh.
Abstract
Radiation hazards to operators performing retrograde endoscopic cholangio-pancreatographies (RECP) have been investigated. The principles and central elements of French medical radioprotection legislation are recalled. Doses received by the operators, over a one month period representative of their usual work, were measured in two ways: by using a ionisation chamber and with thermoluminescent detectors worn by the staff. Extrapolated to 12 months, the results were inferior to the annual reglementary dose limit: whole body < 1.2 mSv (against 50 mSv), eye lens = 38.4 mSv (against 150 mSv). Further means for reducing dose levels are proposed.Mesh:
Year: 1993 PMID: 8339884
Source DB: PubMed Journal: Gastroenterol Clin Biol ISSN: 0399-8320