Abstract

An in vitro comparison of six methods of cerclage wire fixation was performed using 1-mm-diameter stainless steel wire. A testing apparatus was designed to simulate the tensile forces on a cerclage wire around a femoral shaft fracture. The hairpin cerclage knot that we describe generated significantly higher initial compressive force (p < 0.001) and had significantly higher tensile strength (p < 0.02) compared with the other five methods.

Entities:  

Mesh:

Year:  1993        PMID: 8326425     DOI: 10.1097/00005131-199306000-00004

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512