Literature DB >> 8325695

Automatic, noninvasive intradialytic clearance measurement.

H D Polaschegg1.   

Abstract

The feasibility of non invasive, automatic intradialytic measurement of effective clearance is demonstrated by a mathematical derivation. In a practical setting the conductivity difference between dialysate inlet and outlet of the dialyzer is measured for two different dialysate inlet concentrations. This allows for the calculation of the electrolyte dialysance, which is equal to the effective urea clearance, provided dialysate flow, blood flow and blood electrolyte composition remain unchanged during the measuring interval.

Entities:  

Mesh:

Year:  1993        PMID: 8325695

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  8 in total

1.  On-line monitoring of solutes in dialysate using wavelength-dependent absorption of ultraviolet radiation.

Authors:  I Fridolin; L G Lindberg
Journal:  Med Biol Eng Comput       Date:  2003-05       Impact factor: 2.602

Review 2.  On-line dialyser clearance using conductivity.

Authors:  H D Polaschegg
Journal:  Pediatr Nephrol       Date:  1995       Impact factor: 3.714

3.  What volume to choose to assess online Kt/V?

Authors:  Francesco Gaetano Casino; Elena Mancini; Giovanni Santarsia; Salvatore Domenico Mostacci; Filomena D'Elia; Maria Di Carlo; Francesco Iannuzzella; Luigi Rossi; Luigi Vernaglione; Daniela Grimaldi; Renato Rapanà; Carlo Basile
Journal:  J Nephrol       Date:  2019-08-07       Impact factor: 3.902

4.  Evaluation of hemodialysis adequacy using online Kt/V and single-pool variable-volume urea Kt/V.

Authors:  Alicja E Grzegorzewska; Wojciech Banachowicz
Journal:  Int Urol Nephrol       Date:  2008-07-01       Impact factor: 2.370

5.  Measuring plasma conductivity to detect sodium load in hemodialysis patients.

Authors:  Lucile Mercadal; Aude Servais; Marcia Venditto; Nathalie Renault; Corinne Isnard-Bagnis; Gilbert Deray; Thierry Petitclerc
Journal:  Clin J Am Soc Nephrol       Date:  2008-03-27       Impact factor: 8.237

6.  Do only small uremic toxins, chromophores, contribute to the online dialysis dose monitoring by UV absorbance?

Authors:  Jürgen Arund; Risto Tanner; Fredrik Uhlin; Ivo Fridolin
Journal:  Toxins (Basel)       Date:  2012-10-18       Impact factor: 4.546

7.  Interventions to improve hemodialysis adequacy: protocols based on real-time monitoring of dialysate solute clearance.

Authors:  Edward A Ross; Jennifer L Paugh-Miller; Robert W Nappo
Journal:  Clin Kidney J       Date:  2017-10-25

Review 8.  Optimizing haemodialysate composition.

Authors:  Francesco Locatelli; Vincenzo La Milia; Leano Violo; Lucia Del Vecchio; Salvatore Di Filippo
Journal:  Clin Kidney J       Date:  2015-08-08
  8 in total