P Schmalbrock1, M A Brogan, D W Chakeres, V A Hacker, K Ying, B D Clymer.
Abstract
Submillimeter-resolution magnetic resonance (MR) imaging of the inner ear is valuable for diagnosis and treatment planning. Its main advantage for investigations of underlying disease is that it can directly depict the fluid spaces of the membranous labyrinth rather than define only the bony canal, as does computed tomography. A systematic evaluation of factors influencing high-resolution three-dimensional (3D) gradient-echo imaging of the inner ear with a standard clinical MR system is presented. This includes the evaluation of various radio-frequency coils, the design of steady-state pulse sequences, and the optimization of acquisition parameters. A quantitative analysis was facilitated by computer simulations and image processing. The highest signal-to-noise ratio for the membranous labyrinth was obtained with a single 3-inch (7.6-cm) receiver coil and a 3D GRASS (gradient-recalled acquisition in the steady state) sequence with the minimal achievable TR msec/TE msec of 25/7 and a 40 degrees--60 degrees flip angle, which yielded acceptable images with minimal voxel volumes of 0.1 mm3 in 14 minutes.Mesh: