Analytical evaluation of Kone Microlyte determination of ionized magnesium.

H E van Ingen1, H J Huijgen, W T Kok, G T Sanders.   

Abstract

We performed an analytical evaluation of a commercially available instrument for determining ionized magnesium through use of a neutral carrier, liquid-membrane-based ion-selective electrode. Reproducibility (CV 2-4%), linearity (0.30-2.50 mmol/L), lower limit of detection (0.30 mmol/L), and absence of interference from Ca2+ indicate adequate performance for measuring ionized magnesium in plasma or serum samples in the normal to high-concentration range. Sodium in excess of 150 mmol/L caused a negative bias, which can be explained by ionic strength-induced changes in activity coefficients. The use of heparin as an anticoagulant must be restricted to concentrations < 15 units/mL because of the binding of magnesium to heparin. The mean +/- SD concentration of ionized magnesium and its fraction of total magnesium in 76 healthy volunteers were 0.56 +/- 0.05 mmol/L and 0.65 +/- 0.04, respectively.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8287544

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  1 in total

1.  Liquid heparin anticoagulant produces more negative bias in the determination of ionized magnesium than ionized calcium.