Three dimensional shape reconstruction and finite element analysis of femur before and after the cementless type of total hip replacement.

Y K Kang1, H C Park, Y Youm, I K Lee, M H Ahn, J C Ihn.   

Abstract

Computerized tomography was used to reconstruct a shape, and stresses in three-dimensional objects were analysed. The human femur, which has a very irregular shape, was chosen as an object. CT image data of a cadaver femur were transferred to a computer, and an edge extraction program generated the cross-section of bone by specifying a range of CT values for each slice. Pixel data from the CT scan are converted into a vector of points (x, y, z) which can specify the boundaries of bone. Lateral surfaces are defined by stacking up the slices and making use of the vectorized data. Intermediate and oblique cross-sections can be obtained by an interpolation technique. The constructed model was used as input data for the finite element analysis. To understand the stress distributions before and after the cementless type of total hip replacement, a three-dimensional finite element stress analysis of the bone-implant system was carried out, assuming micromotions between the stem and the femur. The analysis was done for both frictionless and friction cases, modelling the contact point with a gap element having isotropic friction. The analysis shows that the stress is not concentrated on the femoral calcar when the friction coefficient is large.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8277755     DOI: 10.1016/0141-5425(93)90065-7

Source DB:  PubMed          Journal:  J Biomed Eng        ISSN: 0141-5425


  3 in total

1.  Creation of the digital three-dimensional model of the prostate and its adjacent structures based on Chinese visible human.

Authors:  Yi Wu; Shao-Xiang Zhang; Na Luo; Ming-Guo Qiu; Li-Wen Tan; Qi-Yu Li; Guang-Jiu Liu; Kai Li
Journal:  Surg Radiol Anat       Date:  2010-04-16       Impact factor: 1.246

2.  Mathematical modeling of fixation of a bone fragment in a new Double-needle external Fixator compared to hoffmann ii fixator.

Authors:  M Pitkin; Y Shukeylo; A Gritsanov
Journal:  Ser Biomech       Date:  2007-12

3.  Mathematical modeling and mechanical and histopathological testing of porous prosthetic pylon for direct skeletal attachment.

Authors:  Mark Pitkin; Grigory Raykhtsaum; John Pilling; Yuri Shukeylo; Vladimir Moxson; Volodimir Duz; John Lewandowski; Raymond Connolly; Robert S Kistenberg; John F Dalton; Boris Prilutsky; Stewart Jacobson
Journal:  J Rehabil Res Dev       Date:  2009
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.