| Literature DB >> 8263097 |
P Houzé1, J Chaussard, P Harry, M Pays.
Abstract
A method has been developed for the separation and measurement of ethylene glycol and three other glycols (propylene glycol, 1,3-butylene glycol and 2,3-butylene glycol) in biological samples by wide-bore column gas chromatography with a flame ionization detector. The method used 1,3-propylene glycol (1,3-propanediol) as an internal standard. The method was linear at least from 2 to 1000 micrograms/ml, with a detection limit of 1 microgram/ml. Analytical recoveries were 89-98% for the different concentrations. Precision studies showed coefficients of variation of 1.5-7.7% for the different concentrations. The assay was applied to the analysis of biological samples from two patients who had ingested ethylene glycol and/or other glycols in a suicide attempt.Entities:
Mesh:
Substances:
Year: 1993 PMID: 8263097 DOI: 10.1016/0378-4347(93)80114-j
Source DB: PubMed Journal: J Chromatogr