| Literature DB >> 8247487 |
M G Harris1, L Fluss, A Lem, H Leong.
Abstract
To evaluate the efficacy of ultraviolet (UV) radiation as a method of disinfecting contact lenses and their storage solutions, we contaminated soft lenses (Bausch & Lomb Optima 38), rigid gas permeable (RGP) lenses (Oxyflow F-30), and their storage solutions with three common bacteria. Escherichia coli (E.c.), Staphylococcus epidermis (S.e.), and Serratia marcescens (S.m.). The storage solutions used were saline solution and RGP conditioning solution. We determined the exposure times to 253.7-nm wavelength UV radiation necessary to disinfect the contact lenses and solutions. The decimal reduction values (D values) found for UV radiation were 10 to 200 hundred times shorter than reported for currently available disinfection systems. For E.c., sterilization was attained after 100 s of exposure. For S.e. and S.m., sterilization occurred after 300 s of exposure. Different contact lens solutions transmit UV radiation to various degrees, with saline solution passing more than 90% of the UV radiation. Thus, our results indicate that UV radiation is an effective and rapid method of disinfecting contact lenses and their storage solutions.Entities:
Mesh:
Year: 1993 PMID: 8247487 DOI: 10.1097/00006324-199310000-00011
Source DB: PubMed Journal: Optom Vis Sci ISSN: 1040-5488 Impact factor: 1.973