High assembly strains and femoral fractures produced during insertion of uncemented femoral components. A cadaver study.

M Jasty1, R M Henshaw, D O O'Connor, W H Harris.   

Abstract

The assembly strains produced in cadaver femurs during uncemented femoral arthroplasty were measured using strain gages and photoelastic coatings. Resecting the femoral neck, reaming the canal with power drills, and rasping with an optimal size rasp, as determined by preoperative radiographic templating, produced small assembly strains, up to 300 microstrain. Insertion of an optimal-size prosthesis after preparing the femoral canal with instruments the same size as the prosthesis produced moderate assembly strains, up to 1,000 microstrain. Half a millimeter press-fit of optimal prostheses produced larger assembly strains, up to 2,000 microstrain. Half a millimeter press-fit of a prosthesis that was also one size (1.0 mm) larger than that determined to be optimum produced even larger assembly strains (2,000-6,000 microstrain) and longitudinal linear fractures in the femoral cortex. Insertion of prostheses that were smaller than the rasps produced minimal strains in the femoral cortex. The magnitude of peak strains produced by press-fitting the femoral components and the small amounts of disparity between the size of the recess and the prosthesis necessary to produce these strains show the narrow range of tolerances available to the surgeon. Cementless femoral arthroplasty requires great care in preparing the femoral canal to the appropriate size as determined from preoperative templating, using accurate and precisely toleranced instrumentation and prosthetic components in order to avoid femoral fractures, yet obtain a stable fit.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8245993

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  4 in total

1.  Digital versus conventional templating techniques in preoperative planning for total hip arthroplasty.

Authors:  Yona Kosashvili; Nadav Shasha; Eli Olschewski; Oleg Safir; Larry White; Allan Gross; David Backstein
Journal:  Can J Surg       Date:  2009-02       Impact factor: 2.089

2.  Significance of proximal femoral computed tomography scanning in the prediction of femoral prosthesis before total hip arthroplasty.

Authors:  Yizhong Li; Jinkuang Lin; Jing Zhang; Jianlong Li; Xuedong Yao; Huafeng Zhuang; Yanjun Zeng
Journal:  Eur J Orthop Surg Traumatol       Date:  2012-01-29

3.  Stable fixation of short-stem femoral implants in patients 70 years and older.

Authors:  Ronak M Patel; Matthew C Smith; Chase C Woodward; S David Stulberg
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

Review 4.  [Endoprosthetic treatment of osteoporosis-related coxarthrosis : aspects of safe patient treatment].

Authors:  S Kirschner; A Hartmann; K-P Günther; C Hamann
Journal:  Orthopade       Date:  2014-04       Impact factor: 1.087

  4 in total