On-line portal imaging: image quality defining parameters for pelvic fields--a clinical evaluation.

D Verellen1, W De Neve, F Van den Heuvel, M Coghe, O Louis, G Storme.   

Abstract

PURPOSE: A test of several image enhancement techniques, performed on on-line portal images in real clinical circumstances, is presented. In addition a score system enabling us to evaluate image quality on pelvic fields is proposed and validated. METHODS AND MATERIALS: Localization images (n = 546) generated by an on-line portal imaging system during the treatment of 13 patients on pelvic fields were obtained by delivering a radiation dose of 6-8 cGy by an 18 MV photon beam, and recorded with a silicon intensified target video camera with adjustable gain, kV- and black level. Set-up errors were corrected before continuing irradiation. A scoring system based on the number of visible bone-soft tissue edges and transformed to a scale 0 to 5 was developed to judge image quality. A validation of this classification of images was performed with the use of transsectional bone-densities (bone-density*radiological path length) specified at the score defining landmarks. A high pass filter was used on all images, additional on-line open field subtraction was performed on 242 fields. Off-line study was performed in which a panel consisting of two groups (one composed of three radiation oncologists, the other of three radiotherapy technologists), scored 470 pelvic fields without further enhancement, and the same images with Contrast Limited Adaptive Histogram Equalization (CLAHE) (Pizer et al.). Two different clipping levels (3.0 and 5.0) were studied.
RESULTS: Gender and transsectional bone-densities were the most defining patient-related factors influencing image quality. Camera settings, gantry angle, and image post-processing were important non-patient-related factors. All investigators judged CLAHE to ameliorate low contrast images and to deteriorate good quality images (p < 0.001).

Entities:  

Mesh:

Year:  1993        PMID: 8244827     DOI: 10.1016/0360-3016(93)90472-8

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  2 in total

1.  Automatic x-ray image contrast enhancement based on parameter auto-optimization.

Authors:  Jianfeng Qiu; H Harold Li; Tiezhi Zhang; Fangfang Ma; Deshan Yang
Journal:  J Appl Clin Med Phys       Date:  2017-09-06       Impact factor: 2.102

2.  Doses delivered by portal imaging quality assurance in routine practice of adjuvant breast radiotherapy worth to by monitored and compensated in some cases.

Authors:  Sami Kefs; Jean-Yves Giraud; Julie Naud; Isabelle Henry; Isabelle Gabelle-Flandin; Jacques Balosso; Abdulhamid Chaikh; Camille Verry
Journal:  Quant Imaging Med Surg       Date:  2021-08
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.