An outbreak of gram-negative bacteremia traced to contaminated O-rings in reprocessed dialyzers.

J P Flaherty1, S Garcia-Houchins, R Chudy, P M Arnow.   

Abstract

OBJECTIVE: To investigate an outbreak of gram-negative bacteremia in an outpatient hemodialysis unit and to identify the source of contaminating bacteria and the route by which bacteria gained access to the bloodstream.
DESIGN: A matched-pair, case-control study and a bacteriologic investigation of the hemodialysis unit and the implicated dialyzers.
SETTING: A university outpatient hemodialysis unit. PATIENTS: Eleven patients receiving long-term hemodialysis who had a total of 12 episodes of primary gram-negative bacteremia and 12 matched controls. MEASUREMENTS: Clinical and demographic data were obtained for patients and controls. Dialysis unit procedures were observed for compliance with aseptic technique. Cultures of potential environmental sources of bacteria were obtained. Hemodialyzers from bacteremic and nonbacteremic patients were dismantled, and the component parts were cultured. Inoculation of O-rings (from Hemoflow F-80 dialyzer) with bacteria and simulated dialysis were done.
RESULTS: During January to October 1988, 12 episodes of primary gram-negative bacteremia caused by Pseudomonas cepacia, Xanthomonas maltophilia, Citrobacter freundii, Acinetobacter calcoaceticus var. anitratus, or Enterobacter cloacae occurred in 11 patients. In 11 episodes, symptoms developed within 3 hours of starting hemodialysis. Intravenous antibiotics were administered for 11 episodes, 3 episodes resulted in hospitalization, and all patients recovered. Case patients were more likely to have received high-flux dialysis with Hemoflow F-80 dialyzers (odds ratio congruent to 11) than were controls. O-rings from dialyzers used by bacteremic patients were culture positive for the organism responsible for bacteremia. Three of the four dialyzers were disinfected using the standard automated method and were recultured 72 hours later; the O-rings of all three dialyzers remained culture positive. Simulated dialysis using dialyzers with contaminated O-rings caused blood pathway contamination despite intervening reprocessing. When the disinfection method for F-80 dialyzers included removal and complete disinfection of the O-rings, O-ring and blood pathway cultures were consistently negative. After this procedure was made routine, no episodes of primary gram-negative bacteremia occurred during the next 6 months.
CONCLUSIONS: Because dialyzers with removable headers and O-rings are widely used in patients receiving long-term hemodialysis, disinfection procedures should include measures to ensure adequate disinfection of O-rings.

Entities:  

Mesh:

Year:  1993        PMID: 8239225     DOI: 10.7326/0003-4819-119-11-199312010-00003

Source DB:  PubMed          Journal:  Ann Intern Med        ISSN: 0003-4819            Impact factor:   25.391


  10 in total

Review 1.  Enterobacter spp.: pathogens poised to flourish at the turn of the century.

Authors:  W E Sanders; C C Sanders
Journal:  Clin Microbiol Rev       Date:  1997-04       Impact factor: 26.132

2.  Spontaneous septic arthritis caused by Burkholderia cepacia.

Authors:  Roberto Augusto Miki; Lee Eric Rubin; Jessica Kirk; Seth D Dodds
Journal:  Iowa Orthop J       Date:  2006

3.  Use of random amplified polymorphic DNA PCR to examine epidemiology of Stenotrophomonas maltophilia and Achromobacter (Alcaligenes) xylosoxidans from patients with cystic fibrosis.

Authors:  J W Krzewinski; C D Nguyen; J M Foster; J L Burns
Journal:  J Clin Microbiol       Date:  2001-10       Impact factor: 5.948

Review 4.  Microbiological and clinical aspects of infection associated with Stenotrophomonas maltophilia.

Authors:  M Denton; K G Kerr
Journal:  Clin Microbiol Rev       Date:  1998-01       Impact factor: 26.132

5.  Hemodialyzer Reuse and Gram-Negative Bloodstream Infections.

Authors:  Chris Edens; Jacklyn Wong; Meghan Lyman; Kyle Rizzo; Duc Nguyen; Michela Blain; Sam Horwich-Scholefield; Heather Moulton-Meissner; Erin Epson; Jon Rosenberg; Priti R Patel
Journal:  Am J Kidney Dis       Date:  2016-12-07       Impact factor: 8.860

6.  Molecular epidemiology of Stenotrophomonas maltophilia isolated from clinical specimens from patients with cystic fibrosis and associated environmental samples.

Authors:  M Denton; N J Todd; K G Kerr; P M Hawkey; J M Littlewood
Journal:  J Clin Microbiol       Date:  1998-07       Impact factor: 5.948

Review 7.  Endotoxemia: methods of detection and clinical correlates.

Authors:  J C Hurley
Journal:  Clin Microbiol Rev       Date:  1995-04       Impact factor: 26.132

8.  Hemodialysis catheter-related bacteremia caused by Stenotrophomonas maltophilia.

Authors:  A Kataria; S Lata; V Khillan
Journal:  Indian J Nephrol       Date:  2015 Sep-Oct

9.  Carbapenem Resistance among Enterobacter Species in a Tertiary Care Hospital in Central India.

Authors:  Atul Khajuria; Ashok Kumar Praharaj; Mahadevan Kumar; Naveen Grover
Journal:  Chemother Res Pract       Date:  2014-08-10

10.  Dialyzer reuse: is it safe and worth it?

Authors:  Ashish Upadhyay
Journal:  J Bras Nefrol       Date:  2019-09-02
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.