The mosaic electrical characteristics of the skin.

D Panescu1, K P Cohen, J G Webster, R A Stratbucker.   

Abstract

The goals of this study are: 1) to characterize the structure of the skin by measuring impedance variations with a suction microelectrode; 2) to correlate the observed impedance variations with dc current pathways through the skin; 3) to characterize the breakdown phenomenon. We constructed a suction microelectrode with a 200-microns internal diameter and performed several tests on two male subjects. Skin impedance measured from different locations on the forearm and palm varied considerably. We found that the average skin impedance on the forearm was larger than the average impedance on the palm and that the ratio between the maximal and minimal skin impedance is larger for the forearm than for the palm. For both the forearm and the palm the magnitude and variance of skin impedance decrease with increasing stimulus frequency. The density of low impedance points observed on the forearm and palm are consistent with the density of dc current pathways through the skin as indicated by traces left on 1-cm2 Ag electrodes when we passed dc current through the skin. The ratio between the highest and lowest impedances decreased as temperature decreased--at low temperatures the skin displayed mostly high impedances. We were not able to break down the skin using the suction microelectrode. The tests with dry and wet electrodes suggest that breakdown is of thermal nature, and that the thermal capacitance of the saline in the suction microelectrode prevents the temperature of the underlying skin from increasing very rapidly. In conjunction with the larger impedance values, this would tend to increase the breakdown voltage.

Entities:  

Mesh:

Year:  1993        PMID: 8225332     DOI: 10.1109/10.243418

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  2 in total

1.  Migration pathways of hypodermically injected technetium-99m in dogs.

Authors:  F M Kovacs; A García; N Mufraggi; F García; J Pavía; D Prandi; V Gotzens; I Giralt; C Piera; J Setoain
Journal:  Eur Radiol       Date:  2000       Impact factor: 5.315

2.  Optimisation of transcutaneous cardiac pacing by three-dimensional finite element modelling of the human thorax.

Authors:  D Panescu; J G Webster; W J Tompkins; R A Stratbucker
Journal:  Med Biol Eng Comput       Date:  1995-11       Impact factor: 2.602

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.