HemoCue beta-glucose photometer evaluated for use in a neonatal intensive care unit.

E Vadasdi1, E Jacobs.   

Abstract

We evaluated the HemoCue beta-Glucose Photometer system for use in our neonatal intensive care unit by assaying 178 heparinized whole-blood samples obtained by heel stick. The required sample size is 5 microL. Plasma glucose was analysed by a glucose oxidase/oxygen electrode methodology. Across the glucose range of 1.28-21.87 mmol/L, the regression slope was 0.976 (r = 0.976, Sy/x = 0.475). For samples with hematocrit < or = 0.30, the regression slope was 0.981 (r = 0.950, Sy/x = 0.415); for hematocrit of 0.31-0.49, the regression slope was 0.984 (r = 0.972, Sy/x = 0.508); and for hematocrit > or = 0.50, the regression slope was 0.959 (r = 0.988, Sy/x = 0.394). Human whole blood, bovine whole blood, and bovine serum-based quality-control materials were studied. Except for assays of the low-concentration human control material, the total CV was < 3.5%. The accuracy and precision of the HemoCue system were comparable with those of conventional laboratory instrumentation.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8222230

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  2 in total

Review 1.  Hypoglycaemia of the newborn: a review.

Authors:  A F Williams
Journal:  Bull World Health Organ       Date:  1997       Impact factor: 9.408

2.  Comparison of two cotside methods for the detection of hypoglycaemia among neonates in Nepal.

Authors:  M Ellis; D S Manandhar; N Manandhar; J M Land; N Patel; A M de L Costello
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1996-09       Impact factor: 5.747

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.