Signal fidelity requirements for deriving impedance cardiographic measures of cardiac function over a broad heart rate range.

B E Hurwitz1, L Y Shyu, C C Lu, S P Reddy, N Schneiderman, J H Nagel.   

Abstract

Our findings indicate that the impedance cardiogram spectrum extends from DC to 50 Hz. Any amplifier with an upper band limit less than 50 Hz can be expected to produce attenuation and distortion of the impedance cardiogram. This signal attenuation may be systematically enhanced under conditions of high heart rate when a greater proportion of signal energy will be in the upper frequency range of the impedance cardiogram spectrum. Therefore, the present study was designed to assess the influence of amplifier bandwidth on dZ/dtmax, stroke volume, and systolic time intervals (LVET, PEP, QZ, QX). Simultaneously measured delta Z and dZ/dt signals from two impedance cardiographs, with corner frequencies of 120 and 60 Hz for the delta Z and 50 and 15 Hz for dZ/dt channels, were contrasted over a broad range of heart rate (70-150 bpm). In addition to the analog dZ/dt signals obtained from the instruments, the delta Z signals were digitally converted to dZ/dt by off-line digital differentiation with a 50 Hz corner frequency. The results demonstrated that the measurements with the 15 Hz corner frequency, when compared with the 50 Hz corner frequency measurements, systematically attenuated the dZ/dtmax amplitude and stroke volume measurements as heart rate increased. The attenuation of dZ/dtmax and stroke volume ranged from about 13% to 26% as heart rate increased from 70 to 150 bpm. When the upper bandlimit was 50 Hz, the dZ/dt signal had greater resolution of waveform events and produced less prolonged systolic time intervals. The 15 Hz amplifier differentially influenced the B point, Z-peak and X minimum, having no apparent effect on the temporal location of the B point, but delaying the Z-peak about 21.7 ms and the X minimum about 7.4 ms. These findings indicate that impedance cardiographs with insufficient upper bandlimits will differentially influence ICG-derived measurements as heart rate varies.

Entities:  

Mesh:

Year:  1993        PMID: 8218621     DOI: 10.1016/0301-0511(93)90076-k

Source DB:  PubMed          Journal:  Biol Psychol        ISSN: 0301-0511            Impact factor:   3.251


  7 in total

1.  Patterns of immune, neuroendocrine, and cardiovascular stress responses in asymptomatic HIV seropositive and seronegative men.

Authors:  K R Starr; M H Antoni; B E Hurwitz; M S Rodriquez; G Ironson; M A Fletcher; M Kumar; R Patarca; S K Lutgendorf; R E Quillian; N G Klimas; N Schneiderman
Journal:  Int J Behav Med       Date:  1996

2.  Autonomic and cardiovascular function in HIV spectrum disease: early indications of cardiac pathophysiology.

Authors:  K A Brownley; J R Milanovich; S J Motivala; N Schneiderman; L Fillion; J A Graves; N G Klimas; M A Fletcher; B E Hurwitz
Journal:  Clin Auton Res       Date:  2001-10       Impact factor: 4.435

3.  Creative motivation: creative achievement predicts cardiac autonomic markers of effort during divergent thinking.

Authors:  Paul J Silvia; Roger E Beaty; Emily C Nusbaum; Kari M Eddington; Thomas R Kwapil
Journal:  Biol Psychol       Date:  2014-07-23       Impact factor: 3.251

4.  Hemodynamic responses during psychological stress: implications for studying disease processes.

Authors:  A Sherwood; J R Turner
Journal:  Int J Behav Med       Date:  1995

5.  Planning, conducting, and analyzing a psychophysiological experiment on challenge and threat: A comprehensive tutorial.

Authors:  Phillip R Johnston; Alexandra E Volkov; William S Ryan; Spike W S Lee
Journal:  Behav Res Methods       Date:  2022-05-23

6.  Validation of a new impedance cardiography analysis algorithm for clinical classification of stress states.

Authors:  Shafa-At Ali Sheikh; Nil Z Gurel; Shishir Gupta; Ikenna V Chukwu; Oleksiy Levantsevych; Mhmtjamil Alkhalaf; Majd Soudan; Rami Abdulbaki; Ammer Haffar; Gari D Clifford; Omer T Inan; Amit J Shah
Journal:  Psychophysiology       Date:  2022-02-12       Impact factor: 4.348

7.  An open-source automated algorithm for removal of noisy beats for accurate impedance cardiogram analysis.

Authors:  Shafa-At Ali Sheikh; Amit Shah; Oleksiy Levantsevych; Majd Soudan; Jamil Alkhalaf; Ali Bahrami Rad; Omer T Inan; Gari D Clifford
Journal:  Physiol Meas       Date:  2020-08-11       Impact factor: 2.688

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.