Literature DB >> 8205236

Enantioselective high-performance liquid chromatographic method for the determination of methadone in serum using an AGP and a CN column as chiral and analytical column, respectively.

K Kristensen1, H R Angelo, T Blemmer.   

Abstract

A simple and sensitive HPLC method with ultraviolet absorption detection at 200 nm is described for the determination of methadone enantiomers in human serum, using dextropropoxyphene as an internal standard and organic solvent extraction. Separation was performed on two serially coupled columns, CN and Chiral AGP, with a mobile phase consisting of acetonitrile, dimethylocytlamine and phosphate buffer. Using 1.0 ml of serum, 5 nmol/1 of each enantiomer could be determined with an acceptable precision. No interactions from several drugs were observed. The method has been successfully used in a pharmacokinetic study. More than 2500 serum samples have been separated on the same AGP column with acceptable selectivity and resolution.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8205236     DOI: 10.1016/0021-9673(94)80390-0

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  1 in total

1.  Distribution of R- and S-methadone into human milk during multiple, medium to high oral dosing.

Authors:  E J Begg; T J Malpas; L P Hackett; K F Ilett
Journal:  Br J Clin Pharmacol       Date:  2001-12       Impact factor: 4.335

  1 in total