| Literature DB >> 8200210 |
E S Kilpatrick1, M J McLeod, A G Rumley, M Small.
Abstract
A 6-week clinical evaluation of the One Touch II blood glucose meter was performed in a medical ward specializing in diabetic patients and compared to the existing Glucometer II system. Nursing staff trained in both methods performed 267 capillary blood glucose measurements using the two meters. A reference capillary blood glucose was performed simultaneously on 129 of these occasions using a Yellow Springs Instrument analyser. The One Touch II correlated better with the reference (r = 0.97) than the Glucometer II (r = 0.86) and was equally close or closer to the Yellow Springs Instrument on 82% of occasions. A greater proportion of samples measured by the One Touch II were within 15% of the reference value (83% vs 66%). There were also 17 cases where the One Touch II was able to give a measurement which was outwith the analytical range of the Glucometer II. A questionnaire revealed the nursing staff unanimously preferred using the One Touch II to their present meter. It is concluded that in a ward setting the One Touch II meter showed improved accuracy, usefulness in its wider analytical range, and a greater user preference in comparison to the Glucometer II.Entities:
Mesh:
Substances:
Year: 1994 PMID: 8200210 DOI: 10.1111/j.1464-5491.1994.tb02023.x
Source DB: PubMed Journal: Diabet Med ISSN: 0742-3071 Impact factor: 4.359