Literature DB >> 8200210

A ward comparison between the One Touch II and Glucometer II blood glucose meters.

E S Kilpatrick1, M J McLeod, A G Rumley, M Small.   

Abstract

A 6-week clinical evaluation of the One Touch II blood glucose meter was performed in a medical ward specializing in diabetic patients and compared to the existing Glucometer II system. Nursing staff trained in both methods performed 267 capillary blood glucose measurements using the two meters. A reference capillary blood glucose was performed simultaneously on 129 of these occasions using a Yellow Springs Instrument analyser. The One Touch II correlated better with the reference (r = 0.97) than the Glucometer II (r = 0.86) and was equally close or closer to the Yellow Springs Instrument on 82% of occasions. A greater proportion of samples measured by the One Touch II were within 15% of the reference value (83% vs 66%). There were also 17 cases where the One Touch II was able to give a measurement which was outwith the analytical range of the Glucometer II. A questionnaire revealed the nursing staff unanimously preferred using the One Touch II to their present meter. It is concluded that in a ward setting the One Touch II meter showed improved accuracy, usefulness in its wider analytical range, and a greater user preference in comparison to the Glucometer II.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8200210     DOI: 10.1111/j.1464-5491.1994.tb02023.x

Source DB:  PubMed          Journal:  Diabet Med        ISSN: 0742-3071            Impact factor:   4.359


  1 in total

1.  Performance of point-of-care diagnostics for glucose, lactate, and hemoglobin in the management of severe malaria in a resource-constrained hospital in Uganda.

Authors:  Michael Hawkes; Andrea L Conroy; Robert O Opoka; Sophie Namasopo; W Conrad Liles; Chandy C John; Kevin C Kain
Journal:  Am J Trop Med Hyg       Date:  2014-03-03       Impact factor: 2.345

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.