Technical feasibility and performance studies of a Doppler guide wire for potential neuroendovascular applications.

J C Chaloupka1, F Viñuela, R P Malanum, C Ji, D E Goller, J Robert, G Duckwiler.   

Abstract

PURPOSE: To conduct technical feasibility and performance studies on a new Doppler-tipped, 0.014-inch micro-guide wire for potential neuroendovascular applications.
METHODS: In vivo microcatheterizations of brachiocephalic arteries were performed in two swine using the 0.014-inch Doppler guide wire and a commonly used microcatheter. A standardized, bench-top method of evaluating basic mechanical properties of micro-guide wires was also used to compare the 0.014-inch Doppler guide wire with a commonly used micro-guide wire.
RESULTS: The 0.014-inch Doppler guide wire had similar steerability, tractability, torque control, and distal tip flexibility to the commonly used micro-guide wire in the in vivo simulations. Frequent micro-guide wire exchanges were possible without loss of superselective positioning of the microcatheter. Bench-top testing showed the 0.014-inch Doppler guide wire to have comparable distal tip flexibility and stiffness to the commonly used micro-guide wire.
CONCLUSION: The comparable subjective and objective mechanical properties of the 0.014-inch Doppler guide wire to that of a commonly used micro-guide wire further establishes the possibility of clinical implementation of the device.

Entities:  

Mesh:

Year:  1994        PMID: 8197947      PMCID: PMC8334313     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  1 in total

1.  Immediate evaluation of angioplasty and stenting results in supra-aortic arteries by use of a Doppler-tipped guidewire.

Authors:  Nina Ghosh; Donatella Tampieri; Denis Melancon
Journal:  AJNR Am J Neuroradiol       Date:  2004-08       Impact factor: 3.825

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.