A comparison of stress-induced porosity due to conventional and a modified spinal fixation device.

T H Lim1, V K Goel, J M Winterbottom, B Kessler, J Y Ahn, J K Gwon, J B Park, J N Weinstein.   

Abstract

A hypothesis that device-related osteopenia can be reduced by decreasing the rigidity of a fixation device was tested through a canine study. Polymer washers were interposed between the integral nut and plate of the variable spinal plating (VSP) system to reduce its rigidity. A solid fusion was observed 6 months postoperatively in all of the animals using VSP or modified systems. The stabilized segment using both systems showed similar load-displacement behaviors immediately after surgery and 6 months postoperatively. Although not significant, 5.6 and 1.8% decreases in volumetric density of mineralized bone were found in the stabilized segments due to VSP and modified systems, respectively. The modified system also increased bone growth around screws. The new concept of using polymer washers, to decrease rigidity of the fixation device over time, may reduce device-related osteopenia.

Entities:  

Mesh:

Year:  1994        PMID: 8186583     DOI: 10.1097/00002517-199407010-00001

Source DB:  PubMed          Journal:  J Spinal Disord        ISSN: 0895-0385


  2 in total

1.  Biomechanical evaluation of pedicle screw-based dynamic stabilization devices for the lumbar spine: a systematic review.

Authors:  Cédric Y Barrey; Ravi K Ponnappan; Jason Song; Alexander R Vaccaro
Journal:  SAS J       Date:  2008-12-01

2.  Pedicle-Screw-Based Dynamic Systems and Degenerative Lumbar Diseases: Biomechanical and Clinical Experiences of Dynamic Fusion with Isobar TTL.

Authors:  Cédric Barrey; Gilles Perrin; Sabina Champain
Journal:  ISRN Orthop       Date:  2013-01-21