L Jiang1, M D Handschumacher, M G Hibberd, S C Siu, M E King, A E Weyman, R A Levine.
Abstract
Two-dimensional echocardiographic measures of right ventricular volume are limited by the asymmetric and crescentic shape of that ventricle and the difficulty in obtaining standardized views. We have developed a three-dimensional echocardiographic system that automatically integrates images and positional data and calculates right ventricular volume without the need for geometric assumptions or standardized views and a surfacing algorithm that takes advantage of the full three-dimensional data set. The accuracy of this system was studied and compared with two-dimensional methods in 12 gel-filled excised human right ventricles (18 to 74 ml). Volumes calculated by three-dimensional echocardiography correlated well with actual values (r = 0.99) and agreed more closely with them than did those obtained by two-dimensional methods (p < 0.02).Entities:
Mesh:
Year: 1994 PMID: 8185959 DOI: 10.1016/s0894-7317(14)80120-x
Source DB: PubMed Journal: J Am Soc Echocardiogr ISSN: 0894-7317 Impact factor: 5.251