Measurement of respiratory mechanics in a mechanically ventilated infant lung simulator: effects of variations in the frequency response of the flow measurement system.

M J Turner1, I M MacLeod, A D Rothberg.   

Abstract

The frequency responses of systems used to measure flow and pressure in ventilated infants may differ, and hence affect estimates of resistance and compliance. We estimated resistance and compliance in 16 ventilated mechanical lung models using linear regression while varying the frequency response of the flow measurement system. Lung models comprised combinations of four sections of tubing and four bottles filled with steel wool. The cut-off frequencies of a filter in the flow measurement system were chosen to yield time delays of 0, +/-3, +/-6, and +/-9 ms relative to the pressure signal. When the phase lags in the measurement systems were not equal at 10 Hz, a bias in resistance approximately (relative delay) x (elastance) ensued. The bias in the resistance estimate when resistance is 5 Pa ml-1 s and compliance is 2 ml kPa-1 is approximately 28% per ms of delay mismatch. Time-shifting the flow data to eliminate the phase discrepancy reduced the resistance bias by 85%. The residual resistance bias was assumed to be due to inappropriate amplitude response. Compliance measurements were affected by less than 8% and less than 2% after time correction of the flow data. Pressure and flow signals must be synchronized to within 1 ms at 10 Hz and the amplitude responses of the measurement systems must be adequate for reliable resistance measurement.

Entities:  

Mesh:

Year:  1994        PMID: 8182962     DOI: 10.1007/bf02512478

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  14 in total

1.  THE ACTION OF ADIABATIC EFFECTS ON THE COMPLIANCE OF AN ARTIFICIAL THORAX.

Authors:  D W HILL; V MOORE
Journal:  Br J Anaesth       Date:  1965-01       Impact factor: 9.166

2.  Measurement of ventilation and respiratory mechanics during continuous positive airway pressure (CPAP) treatment in infants.

Authors:  B Andréasson; M Lindroth; N W Svenningsen; B Drefeldt; G I Jönsson; L Niklason; B Jonson
Journal:  Acta Paediatr Scand       Date:  1989-03

Review 3.  Infant lung function: measurement conditions and equipment.

Authors:  J Stocks; C Beardsmore; P Helms
Journal:  Eur Respir J Suppl       Date:  1989-03

4.  Measurement of the frequency response and common-mode gain of neonatal respiratory pressure and flow measurement systems. Part 1: Apparatus.

Authors:  M J Turner; I M Macleod; A D Rothberg
Journal:  Clin Phys Physiol Meas       Date:  1989-08

5.  Calibration of Fleisch and screen pneumotachographs for use with various oxygen concentrations.

Authors:  M J Turner; I M MacLeod; A D Rothberg
Journal:  Med Biol Eng Comput       Date:  1990-03       Impact factor: 2.602

6.  Flow and volume dependence of respiratory mechanical properties studied by forced oscillation.

Authors:  E Oostveen; R Peslin; C Gallina; A Zwart
Journal:  J Appl Physiol (1985)       Date:  1989-12

7.  Evaluation of neonatal pulmonary mechanics and energetics: a two factor least mean square analysis.

Authors:  V K Bhutani; E M Sivieri; S Abbasi; T H Shaffer
Journal:  Pediatr Pulmonol       Date:  1988

8.  A technique for measuring frequency response of pressure, volume, and flow transducers.

Authors:  A C Jackson; A Vinegar
Journal:  J Appl Physiol Respir Environ Exerc Physiol       Date:  1979-08

9.  Effects of exogenous surfactant therapy on dynamic compliance during mechanical breathing in preterm infants with hyaline membrane disease.

Authors:  R J Couser; T B Ferrara; J Ebert; R E Hoekstra; J J Fangman
Journal:  J Pediatr       Date:  1990-01       Impact factor: 4.406

10.  Respiratory system impedance from 4 to 40 Hz in paralyzed intubated infants with respiratory disease.

Authors:  H L Dorkin; A R Stark; J W Werthammer; D J Strieder; J J Fredberg; I D Frantz
Journal:  J Clin Invest       Date:  1983-09       Impact factor: 14.808

View more
  1 in total

1.  Accuracy of deadspace free ventilatory measurements for lung function testing in ventilated newborns: a simulation study.

Authors:  B Foitzik; P Schaller; M Schmidt; G Schmalisch
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.