Literature DB >> 818129

An evaluation of the Lancer Coagulyzer in the measurement of the one-stage prothrombin time.

L M Rowlands, A C Lawrence.   

Abstract

An evaluation of a photoelectric clot timer, the Lancer Coagulyzer, in measuring the one-stage prothrombin time has been carried out. The machine is considered to provide reliable results on correctly taken samples without interference by plasma bilirubin, only rarely failing to detect clotting in turbid plasma using a wide range of thromboplastins. The apparatus is useful where there is a sufficient number of samples to be tested.

Entities:  

Mesh:

Substances:

Year:  1976        PMID: 818129      PMCID: PMC476029          DOI: 10.1136/jcp.29.3.231

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  1 in total

1.  Evaluation of a photoelectric automatic prothrombin analyzer.

Authors:  F R Davey; L Fiske; A Maltby
Journal:  Am J Clin Pathol       Date:  1972-12       Impact factor: 2.493

  1 in total