P Shapiro, F H Attix, L S August, R B Theus, C C Rogers.
Abstract
The displacement correction factor to be used for analysis of fast-neutron dosimetric measurements using air-filled EG and G tissue-equivalent ion chambers in a tissue-equivalent phantom has been investigated using the MANTA neutron radiotherapy beam generated by 35-MeV deuterons on a thick Be target. The displacement correction factor inferred from these measurements is 0.970 for the EG and G IC-17 (1.0-cm3) ion chamber, and is 0.989 for the EG and G IC-18 (0.1-cm3 ion chamber. This multiplicative displacement correction factor has no significant dependence on depth in the phantom or on neutron beam size.Entities:
Mesh:
Year: 1976 PMID: 817124 DOI: 10.1118/1.594208
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071