Dose distributions produced by a shielded vaginal cylinder using a high-activity iridium-192 source.

F M Waterman1, D E Holcomb.   

Abstract

Algorithms commonly used to compute the dose distribution around a 192Ir source do not calculate the dose by summing the primary and scatter dose components. Thus if the scatter component is reduced significantly by the insertion of a shield into a vaginal cylinder, the algorithm may overestimate the dose delivered to the tumor volume. This problem was investigated by measuring the relative dose around a 2.5 cm diameter shielded vaginal cylinder in a water phantom by use of a 0.147 cm3 thimble ionization chamber. Measurements were made with the cylinder unshielded and with 0.8 cm thick 90 degrees, 180 degrees, and 270 degrees tungsten shields inserted. A reduction in dose was observed on the unshielded side of the cylinder which increased with distance from the source and with the angle subtended by the shield. This reduction did not exceed 2% within 1 cm from the surface of the cylinder with any shield. However, the reduction in dose increased linearly to 3%, 5%, and 15% at a distance of 10 cm from the source with the 90 degrees, 180 degrees, and 270 degrees shields, respectively. On the shielded side of the cylinder, the dose at the surface is reduced to about 13% of its value without the shield; however, this percentage increases with distance reaching a maximum 2-4 cm from the cylinder. The maximum values obtained with the 90 degrees, 180 degrees, and 270 degrees shields were 23%, 17%, and 16%, respectively.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8164574     DOI: 10.1118/1.597241

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Film based verification of calculation algorithms used for brachytherapy planning-getting ready for upcoming challenges of MBDCA.

Authors:  Grzegorz Zwierzchowski; Grzegorz Bielęda; Janusz Skowronek; Magdalena Mazur
Journal:  J Contemp Brachytherapy       Date:  2016-08-16

Review 2.  Review of clinical brachytherapy uncertainties: analysis guidelines of GEC-ESTRO and the AAPM.

Authors:  Christian Kirisits; Mark J Rivard; Dimos Baltas; Facundo Ballester; Marisol De Brabandere; Rob van der Laarse; Yury Niatsetski; Panagiotis Papagiannis; Taran Paulsen Hellebust; Jose Perez-Calatayud; Kari Tanderup; Jack L M Venselaar; Frank-André Siebert
Journal:  Radiother Oncol       Date:  2013-11-30       Impact factor: 6.280

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.