The desflurane (Tec 6) vaporizer: design, design considerations and performance evaluation.

R B Weiskopf1, D Sampson, M A Moore.   

Abstract

We have described the design and design considerations of the desflurane Tec 6 "vaporizer" and have tested its performance characteristics. The vaporizer differs from previous vaporizers designed for anaesthesia in that electromechanical rather than mechanical controls accommodate the different physical characteristics of desflurane. This design, while offering perhaps an increased risk of failure (owing to sophisticated electronic components and circuitry), on the other hand offers the decreased likelihood of accidental delivery of very large concentrations of liquid anaesthetic resulting from tilting or overfilling and alarms and warnings not previously incorporated into the design of anaesthetic vaporizers. The output characteristics of the vaporizer are as expected, based on the design: desflurane concentration output in oxygen has accuracy (+/- 15%) which is similar to that of the mechanical vaporizers; output decreases when nitrous oxide is added owing to the lower viscosity, but remains within 20% of the dial setting or 0.5% absolute.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8155456     DOI: 10.1093/bja/72.4.474

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  7 in total

1.  New volatile anaesthetics in cardiovascular anaesthesia: one step forward, two steps back?

Authors:  I R Thomson
Journal:  Can J Anaesth       Date:  1996-09       Impact factor: 5.063

Review 2.  Sevoflurance: approaching the ideal inhalational anesthetic. a pharmacologic, pharmacoeconomic, and clinical review.

Authors:  L Delgado-Herrera; R D Ostroff; S A Rogers
Journal:  CNS Drug Rev       Date:  2001

Review 3.  Sevoflurane. A review of its pharmacodynamic and pharmacokinetic properties and its clinical use in general anaesthesia.

Authors:  S S Patel; K L Goa
Journal:  Drugs       Date:  1996-04       Impact factor: 9.546

Review 4.  Desflurane. A review of its pharmacodynamic and pharmacokinetic properties and its efficacy in general anaesthesia.

Authors:  S S Patel; K L Goa
Journal:  Drugs       Date:  1995-10       Impact factor: 9.546

5.  Desflurane - revisited.

Authors:  Mukul Chandra Kapoor; Mahesh Vakamudi
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2012-01

6.  Minimal flow anesthesia can be initiated early with the use of higher fresh gas flow to facilitate desflurane "Wash-in".

Authors:  Mukul Chandra Kapoor; Ayalasomayajula Sashank; Ashok Vats; Shaloo Garg; Archana Puri
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2019 Oct-Dec

7.  Comparison of bispectral index-guided and fixed-gas concentration techniques in desflurane and remifentanil anesthesia: A randomized controlled trial.

Authors:  Seong Mi Yang; Yoo Sun Jung; Chul-Woo Jung; Won Ho Kim; Soo Bin Yoon; Hyung-Chul Lee
Journal:  PLoS One       Date:  2020-11-05       Impact factor: 3.240

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.