[The quantitative determination of bone mineral content--a system comparison of similarly built computed tomographs].

R Andresen1, S Radmer, D Banzer, D Felsenberg, K J Wolf.   

Abstract

An intercomparison of 4 CT scanners of the same manufacturer was performed. The bone mineral content of 11 lumbar vertebral columns removed directly post mortem was determined in a specially constructed lucite-water phantom. Even devices of the same construction were shown to yield a variation in the quantitative evaluation markedly exceeding the annual physiological mineral loss. As long as scanner adjustment by physical calibration phantoms has not yet been established, a course assessment and therapy control of bone mineral content should always be carried out on the same QCT scanner.

Entities:  

Mesh:

Year:  1994        PMID: 8136480     DOI: 10.1055/s-2008-1032417

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  2 in total

1.  [Insufficiency fractures of the spine in relation to cancellous bone density : An in vitro study].

Authors:  Guido Schröder; Dirk Flachsmeyer; Claus Maximilian Kullen; Julian Ramin Andresen; Marko Schulze; Laura Hiepe; Hans-Christof Schober; Reimer Andresen
Journal:  Orthopadie (Heidelb)       Date:  2022-05-23

2.  A comparison, using X-ray micro-computed tomography, of the architecture of cancellous bone from the cervical, thoracic and lumbar spine using 240 vertebral bodies from 10 body donors.

Authors:  Guido Schröer; Benjamin Jabke; Marko Schulze; Andreas Wree; Heiner Martin; Olga Sahmel; Alexandra Doerell; Claus Maximilian Kullen; Reimer Andresen; Hans-Christof Schober
Journal:  Anat Cell Biol       Date:  2021-03-31
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.