Correlation of end-tidal CO2 measurements to arterial PaCO2 in nonintubated patients.

C W Barton1, E S Wang.   

Abstract

STUDY
OBJECTIVE: To determine the accuracy of end-tidal carbon dioxide levels as a measure of arterial carbon dioxide levels in nonintubated patients presenting to an emergency department for care.
DESIGN: A prospective, cross-sectional analysis.
SETTING: University hospital ED. TYPE OF PARTICIPANT: Nonintubated adult patients presenting to the ED for care of a variety of problems.
INTERVENTIONS: Patients who had arterial blood gas samples taken as part of their ED evaluation were asked to breathe normally through an endotracheal tube adapter or a modified nasal cannula connected to a side port sampling capnometer while a sample for arterial blood gas was drawn from the radial artery. MEASUREMENTS: End-tidal carbon dioxide levels (mm Hg) were recorded at the time of arterial blood gas sampling. The difference between end-tidal carbon dioxide and PaCO2 was tested with the paired t-test at a significance level of .05. The correlation of end-tidal carbon dioxide to PaCO2 was tested in all patients and in subgroups using simple linear regression.
RESULTS: Seventy-six patients were enrolled. In all patients, end-tidal carbon dioxide was 3.5 mm Hg lower than PaCO2 and correlated well with PaCO2 (r2 = .772). In patients with hypocapnia, there was no significant difference between end-tidal carbon dioxide and PaCO2 (P = .17), and the correlation of end-tidal carbon dioxide to PaCO2 was stronger (r2 = .838). In patients with a respiratory or metabolic acidosis, the difference between end-tidal carbon dioxide and PaCO2 was 6 mm Hg (P = .005), but end-tidal carbon dioxide correlated well to PaCO2 (r2 = .899).
CONCLUSION: Measurements of end-tidal carbon dioxide concentrations correlate well with PaCO2 values in nonintubated patients presenting with a variety of conditions to EDs. End-tidal carbon dioxide measurements may be sufficient measures of PaCO2 in selected patients and obviate the need for repeat arterial blood gas determination. Further study is warranted.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8135434     DOI: 10.1016/s0196-0644(94)70078-8

Source DB:  PubMed          Journal:  Ann Emerg Med        ISSN: 0196-0644            Impact factor:   5.721


  21 in total

1.  A novel airway device with tactile sensing capabilities for verifying correct endotracheal tube placement.

Authors:  Pauwel Goethals; Harshu Chaobal; Dominiek Reynaerts; David Schaner
Journal:  J Clin Monit Comput       Date:  2013-09-25       Impact factor: 2.502

2.  Normal variations in brain oxygen extraction fraction are partly attributed to differences in end-tidal CO2.

Authors:  Dengrong Jiang; Zixuan Lin; Peiying Liu; Sandeepa Sur; Cuimei Xu; Kaisha Hazel; George Pottanat; Sevil Yasar; Paul Rosenberg; Marilyn Albert; Hanzhang Lu
Journal:  J Cereb Blood Flow Metab       Date:  2019-08-05       Impact factor: 6.200

3.  Comparison between end-tidal CO₂ and respiration volume per time for detecting BOLD signal fluctuations during paced hyperventilation.

Authors:  Keith M Vogt; James W Ibinson; Petra Schmalbrock; Robert H Small
Journal:  Magn Reson Imaging       Date:  2011-09-09       Impact factor: 2.546

Review 4.  [Measurement of carbon dioxide in emergency medicine].

Authors:  A Timmermann; J C Brokmann; R Fitzka; E A Nickel
Journal:  Anaesthesist       Date:  2012-02       Impact factor: 1.041

5.  End tidal carbon dioxide as a predictor of the arterial PCO2 in the emergency department setting.

Authors:  C Yosefy; E Hay; Y Nasri; E Magen; L Reisin
Journal:  Emerg Med J       Date:  2004-09       Impact factor: 2.740

6.  Are the local blood oxygen level-dependent (BOLD) signals caused by neural stimulation response dependent on global BOLD signals induced by hypercapnia in the functional MR imaging experiment? Experiments of long-duration hypercapnia and multilevel carbon dioxide concentration.

Authors:  Y J Liu; C J Juan; C Y Chen; C Y Wang; M L Wu; C P Lo; M C Chou; T Y Huang; H Chang; C H Chu; M H Li
Journal:  AJNR Am J Neuroradiol       Date:  2007 Jun-Jul       Impact factor: 3.825

7.  Limitations of End-Tidal CO2 Measured with a Portable Capnometer to Estimate PaCO2 for Patients with Respiratory Disease.

Authors:  Takatoshi Enomoto; Yoshikazu Inoue; Yuichi Adachi; Shunichi Kouno; Yuji Inagaki; Koji Azuma; Kanako Katayama; Naoko Takeuchi; Yoshinobu Matsuda
Journal:  Turk Thorac J       Date:  2021-05

8.  End-tidal CO(2) levels lower in subclinical and overt hypothyroidism than healthy controls; no relationship to thyroid function tests.

Authors:  Khalil Ansarin; Babak Niroomand; Farzad Najafipour; Naser Aghamohammadzadeh; Mitra Niafar; Akbar Sharifi; Mohammadali M Shoja
Journal:  Int J Gen Med       Date:  2011-01-07

Review 9.  State-of-the-art sensor technology in Spain: invasive and non-invasive techniques for monitoring respiratory variables.

Authors:  Christian Domingo; Lluis Blanch; Gaston Murias; Manel Luján
Journal:  Sensors (Basel)       Date:  2010-05-05       Impact factor: 3.576

10.  Clinical commentary: obstetric and respiratory management of pregnancy with severe spinal muscular atrophy.

Authors:  Daniel Flunt; Natasha Andreadis; Collette Menadue; Alec W Welsh
Journal:  Obstet Gynecol Int       Date:  2009-05-19
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.