OBJECTIVE: We propose a simple, compact, and accurate light guidance system for interventional CT. MATERIALS AND METHODS: The system using intersecting laser beams is mounted on the CT gantry and ensures precise needle guidance even when the gantry is tilted. The device does not require additional software/hardware. RESULTS: Phantom simulations and clinical experiments have demonstrated an accuracy of +/- 1 degrees and +/- 1 mm. CONCLUSION: The laser guidance system for CT makes the intervention safer, faster, and more accurate.
OBJECTIVE: We propose a simple, compact, and accurate light guidance system for interventional CT. MATERIALS AND METHODS: The system using intersecting laser beams is mounted on the CT gantry and ensures precise needle guidance even when the gantry is tilted. The device does not require additional software/hardware. RESULTS: Phantom simulations and clinical experiments have demonstrated an accuracy of +/- 1 degrees and +/- 1 mm. CONCLUSION: The laser guidance system for CT makes the intervention safer, faster, and more accurate.
Authors: Rui Li; Sheng Xu; William F Pritchard; John W Karanian; Venkatesh P Krishnasamy; Bradford J Wood; Zion Tsz Ho Tse Journal: Ann Biomed Eng Date: 2018-01-05 Impact factor: 3.934