Anisotropic elasticity and strength of glutaraldehyde fixed bovine pericardium for use in pericardial bioprosthetic valves.

P Zioupos1, J C Barbenel, J Fisher.   

Abstract

Uniaxial tensile tests were performed on glutaraldehyde fixed bovine pericardial strips prepared from chemically modified pericardial samples. These samples originated from an area which demonstrated anisotropic mechanical properties in the native material and which is suitable for the construction of leaflets for pericardial bioprostheses. After glutaraldehyde fixation the tissue had retained its anisotropicity in stiffness and strength in two orthogonal directions. In the range of the functional stresses for a heart valve leaflet (< 1 MPa) the unconstrained fixation regime had modified the initial anisotropic elastic behavior into a more isotropic one. The implications of these findings are that leaflets manufactured from bovine pericardium can be made to resemble, to a degree, the well known anisotropy found in two orthogonal directions in natural human heart valve leaflets, or porcine bioprosthetic heart valve leaflets.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8126028     DOI: 10.1002/jbm.820280107

Source DB:  PubMed          Journal:  J Biomed Mater Res        ISSN: 0021-9304


  7 in total

1.  Comparative study of the mechanical behaviour of a cyanoacrylate and a bioadhesive.

Authors:  J M García Páez; E Jorge Herrero; A Rocha; M Maestro; J L Castillo-Olivares; I Millan; A Carrera Sanmartin; A Cordon
Journal:  J Mater Sci Mater Med       Date:  2004-02       Impact factor: 3.896

2.  Structural mechanism for alteration of collagen gel mechanics by glutaraldehyde crosslinking.

Authors:  Preethi L Chandran; David C Paik; Jeffrey W Holmes
Journal:  Connect Tissue Res       Date:  2012-03-21       Impact factor: 3.417

3.  Patient-specific simulations of transcatheter aortic valve stent implantation.

Authors:  C Capelli; G M Bosi; E Cerri; J Nordmeyer; T Odenwald; P Bonhoeffer; F Migliavacca; A M Taylor; S Schievano
Journal:  Med Biol Eng Comput       Date:  2012-02       Impact factor: 2.602

4.  Evaluation of transcatheter heart valve biomaterials: Biomechanical characterization of bovine and porcine pericardium.

Authors:  Andrés Caballero; Fatiesa Sulejmani; Caitlin Martin; Thuy Pham; Wei Sun
Journal:  J Mech Behav Biomed Mater       Date:  2017-08-09

5.  Mechanical effects of increases in the load applied in uniaxial and biaxial tensile testing: Part I. Calf pericardium.

Authors:  J M García Páez; E Jorge; A Rocha; M Maestro; J L Castillo-Olivares; I Millan; A Carrera; A Cordon; G Tellez; R Burgos
Journal:  J Mater Sci Mater Med       Date:  2002-04       Impact factor: 3.896

6.  Impact of different aortic valve calcification patterns on the outcome of transcatheter aortic valve implantation: A finite element study.

Authors:  Francesco Sturla; Mattia Ronzoni; Mattia Vitali; Annalisa Dimasi; Riccardo Vismara; Georgia Preston-Maher; Gaetano Burriesci; Emiliano Votta; Alberto Redaelli
Journal:  J Biomech       Date:  2016-03-25       Impact factor: 2.712

7.  Biomechanical Identification of High-Risk Patients Requiring Permanent Pacemaker After Transcatheter Aortic Valve Replacement.

Authors:  Guangming Zhang; Rong Liu; Min Pu; Xiaobo Zhou
Journal:  Front Bioeng Biotechnol       Date:  2021-07-09
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.