Comparison of band and spot electrodes for the measurement of stroke volume by the bioelectric impedance technique.

R W Gotshall1, W R Sexson.   

Abstract

OBJECTIVES: Conventional impedance cardiography uses a tetrapolar band electrode configuration to determine thoracic bioelectrical impedance changes and to quantify cardiovascular variables. The present study evaluated the effect of electrode type, band or electrocardiograph (spot), placed in the same anatomical location, on both the absolute value of cardiovascular variables and the percent change in the variable when the patient changed position from supine to standing.
DESIGN: Prospective, randomized studies using repeated-measures design.
SETTING: Research laboratory. PATIENTS: Twenty-one adult volunteer subjects.
INTERVENTIONS: Subjects were studied while supine and again when standing, in random order as to first position.
MEASUREMENTS AND MAIN RESULTS: Stroke volume and impedance-derived variables were measured. Values for stroke volume measured by the band electrode array were 85 +/- 7 and 49 +/- 3 mL for supine and standing, respectively. In contrast, stroke volume measured through the spot electrodes was 110 +/- 8 and 59 +/- 3 mL, respectively. These differences were due to a lower baseline thoracic bioelectrical impedance value measured through the spot electrodes. However, the percent changes for stroke volume with standing subjects were the same when measured through band (-41 +/- 3%) or spot (-44 +/- 3%) electrodes.
CONCLUSIONS: The comparison of cardiovascular values obtained by impedance cardiography should be accomplished only when the values are obtained using similar electrode configuration and type. While there may be a variation in the absolute value of the variable, the direction and magnitude of the change in the variable is relatively unaffected by the electrode types as used in this study.

Entities:  

Mesh:

Year:  1994        PMID: 8124992     DOI: 10.1097/00003246-199403000-00011

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  2 in total

1.  Comparability of Spot Versus Band Electrodes for Impedance Cardiography.

Authors:  Jennifer J McGrath; William H O'Brien; Hilary J Hassinger; Purvi Shah
Journal:  J Psychophysiol       Date:  2005-07-01       Impact factor: 1.333

2.  Planning, conducting, and analyzing a psychophysiological experiment on challenge and threat: A comprehensive tutorial.

Authors:  Phillip R Johnston; Alexandra E Volkov; William S Ryan; Spike W S Lee
Journal:  Behav Res Methods       Date:  2022-05-23
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.