Accuracy of infrared ear thermometry and other temperature methods in adults.

R S Erickson1, L T Meyer.   

Abstract

OBJECTIVE: To compare the accuracy of infrared ear-based temperature measurement in relation to thermometer, ear position, and other temperature methods, with pulmonary artery temperature as the reference.
METHODS: Ear-based temperature measurements were made with four infrared thermometers, three in the core mode and two in the unadjusted mode, each with tug and no-tug techniques. Pulmonary artery, bladder (n = 21), and axillary temperatures were read after each ear-based measurement and oral temperature was measured once when possible (n = 32). Subjects consisted of a convenience sample of 50 patients with pulmonary artery catheters who were in adult critical care units of a university teaching hospital.
RESULTS: Ear-based measurements correlated well with pulmonary artery temperature (r = .87 to .91), although closeness of agreement differed among thermometer-mode combinations (mean offsets = -0.7 to 0.5 degree C) and had moderately high variability between subjects (SD = +/- 0.5 degree C) with all instruments. Use of an ear tug either made no difference or resulted in slightly lower readings. Bladder temperature was nearly identical to pulmonary artery temperature values (r = .99, offset = 0.0 +/- 0.2 degree C). Oral readings were slightly lower (r = .78, offset = -0.2 degree C) and axillary readings much more so (r = .80 to .82, offset = -0.7 degree C); both were highly variable (SD = +/- 0.6 degree C) and affected by external factors.
CONCLUSIONS: Infrared ear thermometry is useful for clinical temperature measurement as long as moderately high variability between patients is acceptable. Readings differ among thermometers, although several instruments provide values close to pulmonary artery temperature in adults. Readings are not higher with an ear tug. Bladder temperature substitutes well for pulmonary artery temperature, whereas oral and axillary values may be influenced by external factors in the critical care setting.

Entities:  

Mesh:

Year:  1994        PMID: 8118492

Source DB:  PubMed          Journal:  Am J Crit Care        ISSN: 1062-3264            Impact factor:   2.228


  11 in total

Review 1.  Thermometry in paediatric practice.

Authors:  A S El-Radhi; W Barry
Journal:  Arch Dis Child       Date:  2006-04       Impact factor: 3.791

2.  The use of infrared ear thermometers in pediatric and family practice offices.

Authors:  B G Silverman; W R Daley; J D Rubin
Journal:  Public Health Rep       Date:  1998 May-Jun       Impact factor: 2.792

Review 3.  Basic principles of optical radiation and some common applications in anesthesia.

Authors:  D Gravenstein; S Lampotang; W Huda; A Sultan
Journal:  J Clin Monit       Date:  1996-11

4.  Temperature measurement in intensive care patients: comparison of urinary bladder, oesophageal, rectal, axillary, and inguinal methods versus pulmonary artery core method.

Authors:  J-Y Lefrant; L Muller; J Emmanuel de La Coussaye; M Benbabaali; C Lebris; N Zeitoun; C Mari; G Saïssi; J Ripart; J-J Eledjam
Journal:  Intensive Care Med       Date:  2003-02-08       Impact factor: 17.440

5.  Comparability of tympanic and oral mercury thermometers at high ambient temperatures.

Authors:  Amy L Chue; Rachael L Moore; Andrew Cavey; Elizabeth A Ashley; Kasia Stepniewska; François Nosten; Rose McGready
Journal:  BMC Res Notes       Date:  2012-07-16

6.  Minimization and management of wastes from biomedical research.

Authors:  E H Rau; R J Alaimo; P C Ashbrook; S M Austin; N Borenstein; M R Evans; H M French; R W Gilpin; J Hughes; S J Hummel; A P Jacobsohn; C Y Lee; S Merkle; T Radzinski; R Sloane; K D Wagner; L E Weaner
Journal:  Environ Health Perspect       Date:  2000-12       Impact factor: 9.031

7.  The effect of humidified heated breathing circuit on core body temperature in perioperative hypothermia during thyroid surgery.

Authors:  Hue Jung Park; Ho Sik Moon; Se Ho Moon; Hyeon Do Jeong; Young Jae Jeon; Keung Do Han; Hyun Jung Koh
Journal:  Int J Med Sci       Date:  2017-07-18       Impact factor: 3.738

8.  Cutaneous infrared thermometry for detecting febrile patients.

Authors:  Pierre Hausfater; Yan Zhao; Stéphanie Defrenne; Pascale Bonnet; Bruno Riou
Journal:  Emerg Infect Dis       Date:  2008-08       Impact factor: 6.883

9.  Reexamining age, race, site, and thermometer type as variables affecting temperature measurement in adults - A comparison study.

Authors:  Linda S Smith
Journal:  BMC Nurs       Date:  2003-06-15

10.  Emergency department rectal temperatures in over 10 years: A retrospective observational study.

Authors:  Graham A Walker; Daniel Runde; Daniel M Rolston; Dan Wiener; Jarone Lee
Journal:  World J Emerg Med       Date:  2013
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.