Literature DB >> 8117364

Stress and strain distribution in the bone surrounding a new design of dental implant: a comparison with a threaded Branemark type implant.

S E Clift1, J Fisher, C J Watson.   

Abstract

The stress and strain distributions in the bone surrounding a new dental implant, designed specifically for use with a bioactive porous coating and thus having a fully bonded interface to the bone, have been analysed. The new implant geometry was slightly tapered, with deep concentric grooves to allow bone ingrowth and load transfer, and had a parallel cylindrical section at the neck. The results have been compared with stress and strain predictions in the bone surrounding a 'Branemark type' threaded implant with a fully bonded interface. Under axial loading both implant types produced similar stress and strain distributions with a higher level of stress in the cortical bone surrounding the neck of the implant. Under lateral loading a high stress concentration was found in the neck region of both implants, but this was lower around the neck of the new design compared with the threaded implant. When the new implant was surrounded by cancellous bone, the reduction in the stress concentration was up to 50 per cent. This reduction should help to reduce fatigue failure and bone resorption in this area under lateral loading.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8117364     DOI: 10.1243/PIME_PROC_1993_207_285_02

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  1 in total

1.  Mathematical modeling and mechanical and histopathological testing of porous prosthetic pylon for direct skeletal attachment.

Authors:  Mark Pitkin; Grigory Raykhtsaum; John Pilling; Yuri Shukeylo; Vladimir Moxson; Volodimir Duz; John Lewandowski; Raymond Connolly; Robert S Kistenberg; John F Dalton; Boris Prilutsky; Stewart Jacobson
Journal:  J Rehabil Res Dev       Date:  2009
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.